Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
Notes supplémentaires dans les données
Cause
The manufacturer states that the referenced devices can have difficulties or failures when deploying the clip, causing prolongation of the procedure time, use of another device and the need for manual compression to achieve the hemostasis, leading to the possible occurrence of adverse events on the patient.
INVIMA 2011RD-0002398 ; INMUNOENSAYO CROMATOGRAFICO RAPIDO PARA LA ; DETECCION CUALITATIVA DE ANTICUERPOS (IgG/IgM) ; PARA TREPONEMA PALLIDUM (TP) EN SANGRE TOTAL, ; SUERO O PLASMA PARA AYUDA DEL DIAGNOSTICO DE ; SIFILIS
Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
Notes supplémentaires dans les données
Cause
Inr results significantly lower than those obtained with a reference system (laboratory method).
Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
Notes supplémentaires dans les données
Cause
They present results of controls outside the established range.
Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
Notes supplémentaires dans les données
Cause
The manufacturer reports that it has detected the possibility that the referenced device provides an inr result significantly lower than that obtained with a laboratory inr system, leading potentially to adverse events on patients.
Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
Notes supplémentaires dans les données
Cause
The manufacturer reports that it has detected that the referenced monitors may provide a result in the prothrombin time (inr) test that is lower than the result obtained from the inr laboratory method, incorrect results may occur if the instructions of the labeling for the completion of the test are not met or by the patient's clinical conditions, leading to the possible generation of adverse events on the patient.
Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
Notes supplémentaires dans les données
Cause
Failures in the use of the aforementioned batches, which cause error codes (code information # 215) when these reagents are stored at room temperature for a shorter period than specified in the insert. according to the insert, storage conditions settle at temperatures of 2-8 ° c, and up to 4 weeks at room temperature (15-25 ° c). the risk to patients is considered low, as it is limited to a potential delay in obtaining a test result.
Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
Notes supplémentaires dans les données
Cause
The fda recently notified the company that it is believed that the company's studies do not adequately demonstrate the effectiveness of the software modification and recommended that alere submit a proposed plan to voluntarily withdraw the inratio device from the market.
Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
Notes supplémentaires dans les données
Cause
The manufacturer states that in the referenced devices a mandrel fracture could occur when the tension is present in the mandrel when turning the actuator knob to unfold the clip, the tension is present if the arm positioner is "closed" in the neutral position, instead to be in the neutral position during the deployment of the clip, leading to potential adverse events being presented to patients.
Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
Notes supplémentaires dans les données
Cause
The manufacturer informs that it has detected that the labeling of the internal packaging does not correspond to the external labeling, leading to the generation of possible delays in the procedures carried out.
Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
Notes supplémentaires dans les données
Cause
Abbott has identified a possible increase in decalibration failures before the expiration date. the reason for this situation is being investigated to find appropriate corrective measures in the long term. it does not affect the results of patients generated. there may be a delay in the generation of results due to calibration errors and calibration failures.
Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
Notes supplémentaires dans les données
Cause
Abbott has confirmed that the drug fulvestrant (faslodex®) can interfere with the architect estradiol trial (reference number: 7k72), which can lead to falsely elevated estradiol results.
Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
Notes supplémentaires dans les données
Cause
Decreased sensitivity that can generate false results.
Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
Notes supplémentaires dans les données
Cause
The manufacturer has determined that some devices manufactured before may 2015 contain batteries that could be discharged sooner than expected "premature battery exhaustion", which could lead to potentially serious adverse events occurring on the patient.
Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
Notes supplémentaires dans les données
Cause
The manufacturer states that it has detected that if the user turns off the system incorrectly they may show an inability to turn on again, leading to delays and possible adverse events on the patients.
Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
Notes supplémentaires dans les données
Cause
It is reported that they may experience a temporary change in their correct functioning when exposed to currents emitted by sources of electrocautery, this change may last thirty (30) seconds or more after the exposure is determined, which may lead to potential adverse events on the patient.
Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
Notes supplémentaires dans les données
Cause
The manufacturer through routine electrical tests has identified faults originating in the manufacturing process, by assembling slightly larger vibrators that could damage the insulation coating of the high voltage capacitor, which would cause a narrowing of the distance between the two adjacent components, resulting in possible leakage of electrical current or an electric arc between the high-voltage capacitor and these components, leading to the occurrence of potential adverse events on patients.