The manufacturer is issuing a recall for in-line blood set with filter, as they have the potential to leak at the tube to cap junction. preliminary investigation has shown that this issue is due to the combination of cap dimension variability, especially on lower limit, and insertion depth of the cap into the tube on the lower side.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Notes supplémentaires dans les données
Recalling Organisation: Medtronic New Zealand Limited, Unit 16 Mezzanine Level, 5 Gloucester Park Road, Onehunga, AUCKLAND
Cause
The fhc depth stop adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. this damage can manifest as low impedances or a short circuit between one or multiple electrode pairs in the lead, and has resulted in the need to remove and replace the implanted lead intraoperatively or in a follow-up procedure.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Notes supplémentaires dans les données
Recalling Organisation: GE Healthcare Limited, 8 Tangihua Street, Auckland 1010
Cause
It has come to the manufacturers attention that the latest on-site software version may not have been reinstalled at some sites after service activities were performed that required software to be reloaded. if the incorrect version is loaded on the system, the images can be flipped left/right and/or there can be patient data mismatch.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Notes supplémentaires dans les données
Recalling Organisation: Philips New Zealand Commercial Limited, Level 3, 123 Carlton Gore Road, Newmarket, AUCKLAND 1023
Cause
The manufacturer has indentified that when the operator or a bystander touches the area of the inner part of the vertical sliding column (on which the tube arm moves up or down), without paying attention of the tube arm movement, the fingers can be pinched.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
This is a reminder to follow the instructions listed in the product documentation when creating a photon beam model in raystation/rayplan. it is the user's responsibility to validate the beam model for all clinically relevant fields before the system is used to create clinical treatment plans.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
The manufacturer has observed that the values for blood and ketones may gradually decrease and values for ph may gradually increase over the shelf life of certain product lots. information has been added to the storage and stability section in the package insert.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Notes supplémentaires dans les données
Recalling Organisation: REM Systems Limited, 69 Elizabeth Knox Place, Glen Innes, AUCKLAND
Cause
It has been identified that there can be a 2mm +/- overall length difference between various manufacturers pp3 batteries that could be used in niki t34 ambulatory syringe pumps., this could result in movement within the battery housing leading to a possible loss of connection, resulting in the pump powering down and a possible under infusion of pain medication occurring.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Notes supplémentaires dans les données
Recalling Organisation: GE Healthcare Limited, 8 Tangihua Street, Auckland 1010
Cause
The manufacturer has identified a small area on the bore surface, which is normally accessible to the patient, can get warm enough to cause a serious burn in the event the patient touches the bore and proper padding is not used.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
The manufacturer has received reports concerning an abnormal coloration of certain agar plates and the growth of non-characteristic mrsa colonies, the origin of the problem seems to be related to certain raw materials. as a result, there is a risk of misinterpretation of the result.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Notes supplémentaires dans les données
Recalling Organisation: GE Healthcare Limited, 8 Tangihua Street, Auckland 1010
Cause
When multiple carescape monitor b650 units are connected to the same network and a network overload occurs for a prolonged time, the monitors may simultaneously restart as designed. the monitor restart will not be completed until the network issue has been corrected. loss of overall monitoring for a prolonged time may lead to a delay in detection of permanent or irreversible impairment or life-threatening changes in the condition of the patient.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
The manufacturer became aware of complaints relating to signs of wear (slight changes in colour and structure) at the handle after the first time of sterilization. whilst the sterilisation has no impact on the electrical safety of the electrodes, it no longer meets the manufacturer's material compatibility acceptance criteria when using the sterrad and steris sterilisation methods.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
The current packaging and ifu indicate a five-year shelf life. during long-term product testing, the manufacturer discovered that after a three-year term, there is a potential for the external packaging (sterile barrier) to become compromised. such compromised packaging results in the loss of the vacuum seal allowing for contamination of the inside peel pouch.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Notes supplémentaires dans les données
Recalling Organisation: Medtronic New Zealand Limited, Unit 16 Mezzanine Level, 5 Gloucester Park Road, Onehunga, AUCKLAND
Cause
The manufacturer has indentified that the irrigation pin used to manufacture (molding) the hub outer tube was incorrectly placed in the mold which resulted in blockage of the irrigation lumen.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Notes supplémentaires dans les données
Recalling Organisation: Johnson & Johnson (NZ) Limited, 13a Gabador Place, Mt Wellington, AUCKLAND
Cause
The manufacturer has identified the potential for debris/material to be found behind the o-rings for some corail neck trials.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Notes supplémentaires dans les données
Recalling Organisation: Getinge Australia Pty Ltd (NZ), 12 G Andromeda Crescent, East Tamaki, Auckland 2013
Cause
The manufacturer has identified that an issue might prevent the device from performing as intended. it was established that, with some of the manufactured units, a screw is missing on the racket of the cupola on some devices in production or that the screws have loosened over the course of life. as a consequence the light head shaft could break and the cupola could detach itself when manipulated and be held only by its cables. it appears such event is more likely to happen mainly during maintenance, cleaning and surgery preparation.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Notes supplémentaires dans les données
Recalling Organisation: Medtronic New Zealand Limited, Unit 16 Mezzanine Level, 5 Gloucester Park Road, Onehunga, AUCKLAND
Cause
This recall is being conducted due to the potential for a device to be misassembled. use of a misassembled device may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Notes supplémentaires dans les données
Recalling Organisation: Medtronic New Zealand Limited, Unit 16 Mezzanine Level, 5 Gloucester Park Road, Onehunga, AUCKLAND
Cause
This recall is being conducted due to the potential for a device to be misassembled. use of a misassembled device may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Notes supplémentaires dans les données
Recalling Organisation: Siemens Healthcare (NZ) Ltd, Millennium Centre, Part Level 2, Building A, 600 Great South Road, Ellerslie, AUCKLAND 1051
Cause
The manufacturer is aware of potential, incorrect tube-current calculations by the care dose4d algorithm when utilizing only pa or ap topograms. depending on the geometrical shape of the skull, rare cases may occur in which the calculated dose distribution is not appropriate for scans that include the head, when based only on pa or ap topograms. the care dose4d software may select the maximum tube current for the uppermost part of the skull, which incidentally leads to unnecessary radiation exposure.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Notes supplémentaires dans les données
Recalling Organisation: Siemens Healthcare (NZ) Ltd, Millennium Centre, Part Level 2, Building A, 600 Great South Road, Ellerslie, AUCKLAND 1051
Cause
The manufacturer is aware of potential, incorrect tube-current calculations by the care dose4d algorithm when utilizing only pa or ap topograms. depending on the geometrical shape of the skull, rare cases may occur in which the calculated dose distribution is not appropriate for scans that include the head, when based only on pa or ap topograms. the care dose4d software may select the maximum tube current for the uppermost part of the skull, which incidentally leads to unnecessary radiation exposure.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Notes supplémentaires dans les données
Recalling Organisation: Siemens Healthcare (NZ) Ltd, Millennium Centre, Part Level 2, Building A, 600 Great South Road, Ellerslie, AUCKLAND 1051
Cause
The manufacturer is aware of potential, incorrect tube-current calculations by the care dose4d algorithm when utilizing only pa or ap topograms. depending on the geometrical shape of the skull, rare cases may occur in which the calculated dose distribution is not appropriate for scans that include the head, when based only on pa or ap topograms. the care dose4d software may select the maximum tube current for the uppermost part of the skull, which incidentally leads to unnecessary radiation exposure.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Notes supplémentaires dans les données
Recalling Organisation: Philips New Zealand Commercial Limited, Level 3, 123 Carlton Gore Road, Newmarket, AUCKLAND 1023
Cause
A problem has been detected in the xper flex cardio patient monitoring system, it is possible for any of the following conditions to occur, which could result in delays in diagnosis or treatment:, 1. boom monitor (display) may not display all active waveform and/or vital sign data., 2. delayed audible and visual alarms at the flex cardio and boom monitor respectively, due to a mismatch between the alarm limits on the host system and the flex cardio/boom monitor., 3. after patient admission and case initiation at the host system, the case may not be fully transmitted to the flex cardio.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Notes supplémentaires dans les données
Recalling Organisation: Smiths Medical Australasia Pty Ltd, c/o KPMG, KPMG Centre, 18 Viaduct Harbour Avenue, Auckland
Cause
The manufacturer has identified that several lots of infusion stopcocks may potentially have punctures in their laminate film packaging. these products are subject to field removal due to the potential for compromised sterility integrity.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
The manufacturer has recently become aware that the abl90 flex in a very rare specific situation can aspirate cal 2 solution instead of rinse solution., the specific situation can occur when the following take place:, 1. the analyzer is in ready mode, 2. operator lifts the inlet and then closes it again, thereby initiating a rinse, 3. immediately after lifts the inlet and quickly closes it again.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Notes supplémentaires dans les données
Recalling Organisation: Philips New Zealand Commercial Limited, Level 3, 123 Carlton Gore Road, Newmarket, AUCKLAND 1023
Cause
The fixation of the upper and lower tilt actuator, of the omnidiagnost system might break off and the table will start to rotate from 0 to + 90 /- 90 degree with high speed. this rotating movement can not be stopped by the user.