Neonatal phototherapy unit - Product Code LBI
Cause
Replacement led board kit of natus neoblue2 phototherapy system distributed after february 16, 2012 is subject of a field safety notification because the system will exhibit a higher light intensity output than the original device. while some customers may prefer using these lights at the higher intensity levels, some are concerned that patients may receive more intensity than desired.
Action
Natus Medical has begun mailing the Urgent Recall Notification/Field Safety Notice neoBLUE LED Board Replacements letter (DOC-012737B), dated November 5, 2014, to US consignees.
Natus Medical also send the Product Advisory neoBLUE LED Board Replacements letter (DOC-0012738), dated October 30, 2014, to International Distribution Partners.
Both US and International consignees also received the TECHNICAL BULLETIN LED Board Kit, neoBLUE 2 LED Phototherapy System, p/n: 001840, Treatment Intensity with revised LED board for neoBLUE 2 systems (p/n: 040904 & 040906) [DOC 008353]; and the Label, neoBLUE Intensity/Height (DOC 011561).
Consignees are instructed to measure the light intensity of their devices and to document that information of the device using an adhesive label that is provided.
The neoBLUE 3 system and its replacement panel sold today are not
affected by this issue. If there are any questions regarding this advisory letter or the attached Technical Bulletin, please call Natus Technical Service at 888-496-2887.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Roche diabetes care was informed by unomedical, the manufacturer of accu-chek tender i and accu-chek tender ii infusion sets, about potential for the
the tubing of the infusion set to become detached at the connect/disconnect location on
the accu-chek tender i and accu-chek tender ii infusion sets. if tubing detachment occurs, insulin delivery is interrupted and the pump will not alarm to notify.
Action
Roche sent the "Urgent Medical Device Corrections" letter dated 12/1/2014 via UPS Ground (receipt required) to the Patients, and Distributors who received the affected device. The letter included a description of the reason for recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. No product returns were expected. For questions, contact the ACCU-CHEK Pump Support at 1-800-688-4578.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Roche diabetes care was informed by unomedical, the manufacturer of accu-chek tender i and accu-chek tender ii infusion sets, about potential for the
the tubing of the infusion set to become detached at the connect/disconnect location on
the accu-chek tender i and accu-chek tender ii infusion sets. if tubing detachment occurs, insulin delivery is interrupted and the pump will not alarm to notify.
Action
Roche sent the "Urgent Medical Device Corrections" letter dated 12/1/2014 via UPS Ground (receipt required) to the Patients, and Distributors who received the affected device. The letter included a description of the reason for recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. No product returns were expected. For questions, contact the ACCU-CHEK Pump Support at 1-800-688-4578.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Roche diabetes care was informed by unomedical, the manufacturer of accu-chek tender i and accu-chek tender ii infusion sets, about potential for the
the tubing of the infusion set to become detached at the connect/disconnect location on
the accu-chek tender i and accu-chek tender ii infusion sets. if tubing detachment occurs, insulin delivery is interrupted and the pump will not alarm to notify.
Action
Roche sent the "Urgent Medical Device Corrections" letter dated 12/1/2014 via UPS Ground (receipt required) to the Patients, and Distributors who received the affected device. The letter included a description of the reason for recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. No product returns were expected. For questions, contact the ACCU-CHEK Pump Support at 1-800-688-4578.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Roche diabetes care was informed by unomedical, the manufacturer of accu-chek tender i and accu-chek tender ii infusion sets, about potential for the
the tubing of the infusion set to become detached at the connect/disconnect location on
the accu-chek tender i and accu-chek tender ii infusion sets. if tubing detachment occurs, insulin delivery is interrupted and the pump will not alarm to notify.
Action
Roche sent the "Urgent Medical Device Corrections" letter dated 12/1/2014 via UPS Ground (receipt required) to the Patients, and Distributors who received the affected device. The letter included a description of the reason for recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. No product returns were expected. For questions, contact the ACCU-CHEK Pump Support at 1-800-688-4578.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Roche diabetes care was informed by unomedical, the manufacturer of accu-chek tender i and accu-chek tender ii infusion sets, about potential for the
the tubing of the infusion set to become detached at the connect/disconnect location on
the accu-chek tender i and accu-chek tender ii infusion sets. if tubing detachment occurs, insulin delivery is interrupted and the pump will not alarm to notify.
Action
Roche sent the "Urgent Medical Device Corrections" letter dated 12/1/2014 via UPS Ground (receipt required) to the Patients, and Distributors who received the affected device. The letter included a description of the reason for recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. No product returns were expected. For questions, contact the ACCU-CHEK Pump Support at 1-800-688-4578.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notes supplémentaires dans les données
Disinfector, medical devices - Product Code MEC
Cause
Getinge disinfection ab received complaints regarding sediment residuals in the manifold of the washer and disinfector of getinge 46-series (cm302) with pacs 300/350 control. after performing detailed root cause analysis, the firm has identified the problem as insufficient water flow throughout the manifold due to formation of air pockets where sedimentation may occur.
Action
An Urgent Field Safety Notice was sent to consignees in North America via USPS Certified Mail with signature receipt on 11/25/14 informing them of the correction. The actions that will be carried out will fully address the problem of sedimentation i.e. stop any future sedimentation. You will be contacted by a Getinge representative to schedule a visit. No additional precautions need to be taken by the end user until the Getinge representative arrives. For questions, please contact Getinge USA at 1-800-320-3532.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notes supplémentaires dans les données
Wrap, sterilization - Product Code FRG
Cause
Advanced sterilization products (asp) is recalling the tyvek rolls with sterrad chemical indicator because they were shipped to customers after its expiration date of may 18, 2013.
Action
A customer notification letter was sent on 12/24/14 to all customers who purchased the expired Tyvek Rolls with STERRAD Chemical Indicator. The letter informs the customers of the problems identified and the actions to be taken.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Carefusion is recalling the smartsite low sorbing infusion set due to disconnection and leakage issues.
Action
CareFusion sent an Urgent Medical Device Recall Notification letter dated January 13, 2015, to all affected customers. The letter informs the customers of the problems identified and the actions to be taken. Customers were instructed to complete and return to CareFusion the recall response card. Customers with questions were instructed to CareFusion: Recall related questions, (888) 562-6018; Adverse Event Reports, (888) 812-3266; Technical Questions regarding the Alaris System, (888) 812-3229.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Shaft breaks so that adequate patient specimen sample cannot be obtained.
Action
Consignees were initially notified of the recall via phone on 12/23/2014 and 12/29/2014. They were informed that the product may have been over exposed to gamma radiation, resulting in brittle sticks. They were instructed to not use the product for patient sampling and to return any product remaining in inventory to Puritan.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notes supplémentaires dans les données
Pump, infusion - Product Code FRN
Cause
Cme america is recalling the bodyguard and bodyguard 323 infusion pumps due to the potential for an over delivery.
Action
CME America sent an "Urgent Medical Device Recall" letter dated January 16, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer.
Required Action for Users: To date there have been no reports of adverse events or serious injury. However, there is the potential for serious injury or death. CME America recommends that you DO NOT use the BodyGuard 323 pump for PCA infusion. If your devices are currently in use, be sure that the programmed parameters DO NOT involve a patient controlled bolus dose less than 0.1 miln volume (translated into meg or mg).
Follow-up Actions by CME America: CME America has identified the root cause of this issue and is recommending installation of a new version of software, USA36B4, for affected BodyGuard 323 infusion pumps. This software is identical to the previously released version, USA36A4.2. However the designation of B4 in the software name will identify updated software in each device. CME America will contact all affected customers within 60 days to initiate the scheduling process for the remediation.
Your assistance is appreciated and necessary to prevent use and facilitate remediation of the recalled product. CME America can address any questions regarding the recalled product by calling the CME America recall line at 720-839-2391.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notes supplémentaires dans les données
Gown, surgical - Product Code FYA
Cause
A single lot (11121171) of flyte toga peel-away x-large (catalog number 0408-830-100) was labeled with an incorrect expiration date of 2412-12-01. the correct expiration date should have read 2014-12-01.
Action
Stryker sent an Urgent Medical Device Recall Notification letter dated December 22, 2014, to all affected customers. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and responding to the formal recall notification. Customers were instructed to do the following:
1. Immediately review this Recall Notification
2. Immediately check all stock areas and/or operating room storage and quarantine any affected product found.
3. Complete the enclosed Business Reply Form (BRF) to confirm receipt of this notification and identify how
many, if any, affected items are currently in their inventory. Complete and return the BRF even if they
do not have any affected product on hand.
Note: Your signature on the BRF indicates that you received and understand this Notification and
have followed the instructions in this Notification.
4. If customers further distributed this product, they were instructed to forward this letter and the attached Business Reply Form (BRF) to all affected locations. Please indicate each location on the BRF. Completed Business Reply Forms should be faxed to 866-521-2762.
Upon receipt of the Business Reply Form, Stryker will email a pre-paid shipper. This shipper can be used to return the recalled Flyte Toga(s) to Stryker.
Upon receipt of the recalled Flyte Toga(s), credit will be applied to the customer's account.
For questions regarding this recall customers were instructed to contact Stryker Instruments: Monday-Friday 8am-5pm (EST) at 269-389-2458.
For questions regarding this recall call 269-389-2458.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Potential for leaking caused by insufficient bond between needleless device and female luer.
Action
Vygon sent an Urgent Medical Device Recall letter dated January 15, 2015, to all affected customers via email or courier service. The letter instructs consignees to check stock for products included in the scope of the recall, count and document the affected inventory on the "Recall Acknowledgement and Inventory Return Form" and fax or email the form to Vygon. Vygon customer service will then contact the customer to issue a Return Merchandise Authorization Number so the inventory can be replaced at no cost. Distributors were provided with instructions on how to handle product that was further distributed by them.
If after 30-days, product is not returned from each consignee, Vygon will follow-up with the consignee to obtain a status of the product return.
Customers with questions were instructed to contact Vygon Customer Service at 1-800-473-5414, or by e-mail at customerservice@vygonus.com.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Potential for leaking caused by insufficient bond between needleless device and female luer.
Action
Vygon sent an Urgent Medical Device Recall letter dated January 15, 2015, to all affected customers via email or courier service. The letter instructs consignees to check stock for products included in the scope of the recall, count and document the affected inventory on the "Recall Acknowledgement and Inventory Return Form" and fax or email the form to Vygon. Vygon customer service will then contact the customer to issue a Return Merchandise Authorization Number so the inventory can be replaced at no cost. Distributors were provided with instructions on how to handle product that was further distributed by them.
If after 30-days, product is not returned from each consignee, Vygon will follow-up with the consignee to obtain a status of the product return.
Customers with questions were instructed to contact Vygon Customer Service at 1-800-473-5414, or by e-mail at customerservice@vygonus.com.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Potential for leaking caused by insufficient bond between needleless device and female luer.
Action
Vygon sent an Urgent Medical Device Recall letter dated January 15, 2015, to all affected customers via email or courier service. The letter instructs consignees to check stock for products included in the scope of the recall, count and document the affected inventory on the "Recall Acknowledgement and Inventory Return Form" and fax or email the form to Vygon. Vygon customer service will then contact the customer to issue a Return Merchandise Authorization Number so the inventory can be replaced at no cost. Distributors were provided with instructions on how to handle product that was further distributed by them.
If after 30-days, product is not returned from each consignee, Vygon will follow-up with the consignee to obtain a status of the product return.
Customers with questions were instructed to contact Vygon Customer Service at 1-800-473-5414, or by e-mail at customerservice@vygonus.com.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notes supplémentaires dans les données
Needle, hypodermic, single lumen - Product Code FMI
Cause
Needle is not captured in the needle safety sheath in specific lots.
Action
Smith Medical sent on 11/24/2014, consignees an Urgent Field Safety Notice via FedEx Overnight. The notification instructs customers to inspect their inventory and complete and return the attached Urgent Field Safety Notice Confirmation Form to coordinate return of affected product to Smiths Medical for replacement/credit. For questions regarding this information, please contact Smiths Medicals Customer Service Department at 800-258-5361.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notes supplémentaires dans les données
Needle, hypodermic, single lumen - Product Code FMI
Cause
Needle is not captured in the needle safety sheath in specific lots.
Action
Smith Medical sent on 11/24/2014, consignees an Urgent Field Safety Notice via FedEx Overnight. The notification instructs customers to inspect their inventory and complete and return the attached Urgent Field Safety Notice Confirmation Form to coordinate return of affected product to Smiths Medical for replacement/credit. For questions regarding this information, please contact Smiths Medicals Customer Service Department at 800-258-5361.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notes supplémentaires dans les données
Needle, hypodermic, single lumen - Product Code FMI
Cause
Needle is not captured in the needle safety sheath in specific lots.
Action
Smith Medical sent on 11/24/2014, consignees an Urgent Field Safety Notice via FedEx Overnight. The notification instructs customers to inspect their inventory and complete and return the attached Urgent Field Safety Notice Confirmation Form to coordinate return of affected product to Smiths Medical for replacement/credit. For questions regarding this information, please contact Smiths Medicals Customer Service Department at 800-258-5361.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notes supplémentaires dans les données
Medical device data system - Product Code OUG
Cause
Horiba medical is informing all horiba medical lite dm customers using software version 2.0.0.19 and lower of a software bug related to the potential for results received being applied to the wrong order in the lite dm worklist. this bug has not been seen in software versions above v2.0.0.19
impact: potential for results received being applied to the wrong order in the lite dm worklist. the is.
Action
ACTION/RESOLUTION:
Before accepting any results please verify the Sample ID in the Worklist matches the Sample ID of the Patient information displayed for the results (Figure 1 below). If the two numbers do not match please contact HORIBA Medical Technical Support.
Your Lite DM should be upgraded to V3.0.0.21. Please contact HORIBA Medical Technical Support to schedule an upgrade. Please call the following phone # 1(888) 903-5001 prompt # 3.
Please share this information with your laboratory staff and retain this notification as part of your laboratory documentation. It is mandatory for you to complete and return the enclosed response form within 10 days, in order for us to verify your
understanding of this Field Information Notice.
If you have any questions or concerns, please contact your local HORIBA Medical representative. We sincerely apologize for any inconvenience this may cause your laboratory.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notes supplémentaires dans les données
Gown, surgical - Product Code FYA
Cause
This recall is being conducted because a cohesive band used to package the devices contains latex and the product is not labeled as required by 21 cfr 801.437; it does not contain caution: the packaging of this product contains natural rubber latex which may cause allergic reactions.
Action
The firm, Ecolab - Microtek Medical, sent an "Urgent: Recall - Mislabeled Product Immediate Action Required" letter dated March 2015 to its US customers on 03/10/2015 via letter and email. The letter describes product, problem and actions to be taken. The customers were instructed to immediately examine your inventory and promptly quarantine the lot(s) listed; complete and return the enclosed Recall Response Form via fax to Quality Assurance Manager, Ecolab-Microtek Medical at (662) 244-3011 or email to: recall@ecolab.com RE: Recall - Mislabeled Product ASAP, and if you have further distributed the products subject to this recall, please notify your customers at once.
Upon receipt of the completed response forms, Ecolab-Microtek Medical will issue a return material authorization (RMA) and ask that you return any unopened product listed above to Ecolab-Microtek Medical at the address listed in the attached form.
If you have any questions in regard to this recall, please contact Customer Service by phone at 1-800-824-3027 or email; customerservice@microtekmed.com.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notes supplémentaires dans les données
Gown, surgical - Product Code FYA
Cause
This recall is being conducted because a cohesive band used to package the devices contains latex and the product is not labeled as required by 21 cfr 801.437; it does not contain caution: the packaging of this product contains natural rubber latex which may cause allergic reactions.
Action
The firm, Ecolab - Microtek Medical, sent an "Urgent: Recall - Mislabeled Product Immediate Action Required" letter dated March 2015 to its US customers on 03/10/2015 via letter and email. The letter describes product, problem and actions to be taken. The customers were instructed to immediately examine your inventory and promptly quarantine the lot(s) listed; complete and return the enclosed Recall Response Form via fax to Quality Assurance Manager, Ecolab-Microtek Medical at (662) 244-3011 or email to: recall@ecolab.com RE: Recall - Mislabeled Product ASAP, and if you have further distributed the products subject to this recall, please notify your customers at once.
Upon receipt of the completed response forms, Ecolab-Microtek Medical will issue a return material authorization (RMA) and ask that you return any unopened product listed above to Ecolab-Microtek Medical at the address listed in the attached form.
If you have any questions in regard to this recall, please contact Customer Service by phone at 1-800-824-3027 or email; customerservice@microtekmed.com.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notes supplémentaires dans les données
Gown, surgical - Product Code FYA
Cause
This recall is being conducted because a cohesive band used to package the devices contains latex and the product is not labeled as required by 21 cfr 801.437; it does not contain caution: the packaging of this product contains natural rubber latex which may cause allergic reactions.
Action
The firm, Ecolab - Microtek Medical, sent an "Urgent: Recall - Mislabeled Product Immediate Action Required" letter dated March 2015 to its US customers on 03/10/2015 via letter and email. The letter describes product, problem and actions to be taken. The customers were instructed to immediately examine your inventory and promptly quarantine the lot(s) listed; complete and return the enclosed Recall Response Form via fax to Quality Assurance Manager, Ecolab-Microtek Medical at (662) 244-3011 or email to: recall@ecolab.com RE: Recall - Mislabeled Product ASAP, and if you have further distributed the products subject to this recall, please notify your customers at once.
Upon receipt of the completed response forms, Ecolab-Microtek Medical will issue a return material authorization (RMA) and ask that you return any unopened product listed above to Ecolab-Microtek Medical at the address listed in the attached form.
If you have any questions in regard to this recall, please contact Customer Service by phone at 1-800-824-3027 or email; customerservice@microtekmed.com.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notes supplémentaires dans les données
Gown, surgical - Product Code FYA
Cause
This recall is being conducted because a cohesive band used to package the devices contains latex and the product is not labeled as required by 21 cfr 801.437; it does not contain caution: the packaging of this product contains natural rubber latex which may cause allergic reactions.
Action
The firm, Ecolab - Microtek Medical, sent an "Urgent: Recall - Mislabeled Product Immediate Action Required" letter dated March 2015 to its US customers on 03/10/2015 via letter and email. The letter describes product, problem and actions to be taken. The customers were instructed to immediately examine your inventory and promptly quarantine the lot(s) listed; complete and return the enclosed Recall Response Form via fax to Quality Assurance Manager, Ecolab-Microtek Medical at (662) 244-3011 or email to: recall@ecolab.com RE: Recall - Mislabeled Product ASAP, and if you have further distributed the products subject to this recall, please notify your customers at once.
Upon receipt of the completed response forms, Ecolab-Microtek Medical will issue a return material authorization (RMA) and ask that you return any unopened product listed above to Ecolab-Microtek Medical at the address listed in the attached form.
If you have any questions in regard to this recall, please contact Customer Service by phone at 1-800-824-3027 or email; customerservice@microtekmed.com.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notes supplémentaires dans les données
Apparatus, exhaust, surgical - Product Code FYD
Cause
This recall is being conducted because a cohesive band used to package the devices contains latex and the product is not labeled as required by 21 cfr 801.437; it does not contain caution: the packaging of this product contains natural rubber latex which may cause allergic reactions.
Action
The firm, Ecolab - Microtek Medical, sent an "Urgent: Recall - Mislabeled Product Immediate Action Required" letter dated March 2015 to its US customers on 03/10/2015 via letter and email. The letter describes product, problem and actions to be taken. The customers were instructed to immediately examine your inventory and promptly quarantine the lot(s) listed; complete and return the enclosed Recall Response Form via fax to Quality Assurance Manager, Ecolab-Microtek Medical at (662) 244-3011 or email to: recall@ecolab.com RE: Recall - Mislabeled Product ASAP, and if you have further distributed the products subject to this recall, please notify your customers at once.
Upon receipt of the completed response forms, Ecolab-Microtek Medical will issue a return material authorization (RMA) and ask that you return any unopened product listed above to Ecolab-Microtek Medical at the address listed in the attached form.
If you have any questions in regard to this recall, please contact Customer Service by phone at 1-800-824-3027 or email; customerservice@microtekmed.com.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notes supplémentaires dans les données
Apparatus, exhaust, surgical - Product Code FYD
Cause
This recall is being conducted because a cohesive band used to package the devices contains latex and the product is not labeled as required by 21 cfr 801.437; it does not contain caution: the packaging of this product contains natural rubber latex which may cause allergic reactions.
Action
The firm, Ecolab - Microtek Medical, sent an "Urgent: Recall - Mislabeled Product Immediate Action Required" letter dated March 2015 to its US customers on 03/10/2015 via letter and email. The letter describes product, problem and actions to be taken. The customers were instructed to immediately examine your inventory and promptly quarantine the lot(s) listed; complete and return the enclosed Recall Response Form via fax to Quality Assurance Manager, Ecolab-Microtek Medical at (662) 244-3011 or email to: recall@ecolab.com RE: Recall - Mislabeled Product ASAP, and if you have further distributed the products subject to this recall, please notify your customers at once.
Upon receipt of the completed response forms, Ecolab-Microtek Medical will issue a return material authorization (RMA) and ask that you return any unopened product listed above to Ecolab-Microtek Medical at the address listed in the attached form.
If you have any questions in regard to this recall, please contact Customer Service by phone at 1-800-824-3027 or email; customerservice@microtekmed.com.