Cause
Device failure / feature before use - material / component.
Finnish data is current through April 2018. All of the data comes from Valvira - National Supervisory Authority for Welfare and Health, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Finland.
Notes supplémentaires dans les données
Source: NCAR
Cause
Device failure / feature after commissioning - mechanical failure.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Fully threaded taper post fixation components not cleared for marketing in the us with current indications.
Action
Arthosurface sent an Urgent Medical Device Recall letter on November 24, 2014 via UPS, to all affected customers. The letter informed consignees of the recall and asked them to return two completed forms, 1. Acknowledgement & Receipt Form and 2. Inventory Record via fax, email, or mail. Customers with questions were instcted to contact their local sales representatiave or email RA@arthrosurface.com. For questions regarding this recall call 508-528-4604.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Fully threaded taper post fixation components not cleared for marketing in the us with current indications.
Action
Arthosurface sent an Urgent Medical Device Recall letter on November 24, 2014 via UPS, to all affected customers. The letter informed consignees of the recall and asked them to return two completed forms, 1. Acknowledgement & Receipt Form and 2. Inventory Record via fax, email, or mail. Customers with questions were instcted to contact their local sales representatiave or email RA@arthrosurface.com. For questions regarding this recall call 508-528-4604.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Fully threaded taper post fixation components not cleared for marketing in the us with current indications.
Action
Arthosurface sent an Urgent Medical Device Recall letter on November 24, 2014 via UPS, to all affected customers. The letter informed consignees of the recall and asked them to return two completed forms, 1. Acknowledgement & Receipt Form and 2. Inventory Record via fax, email, or mail. Customers with questions were instcted to contact their local sales representatiave or email RA@arthrosurface.com. For questions regarding this recall call 508-528-4604.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Fully threaded taper post fixation components not cleared for marketing in the us with current indications.
Action
Arthosurface sent an Urgent Medical Device Recall letter on November 24, 2014 via UPS, to all affected customers. The letter informed consignees of the recall and asked them to return two completed forms, 1. Acknowledgement & Receipt Form and 2. Inventory Record via fax, email, or mail. Customers with questions were instcted to contact their local sales representatiave or email RA@arthrosurface.com. For questions regarding this recall call 508-528-4604.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Fully threaded taper post fixation components not cleared for marketing in the us with current indications.
Action
Arthosurface sent an Urgent Medical Device Recall letter on November 24, 2014 via UPS, to all affected customers. The letter informed consignees of the recall and asked them to return two completed forms, 1. Acknowledgement & Receipt Form and 2. Inventory Record via fax, email, or mail. Customers with questions were instcted to contact their local sales representatiave or email RA@arthrosurface.com. For questions regarding this recall call 508-528-4604.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Analysis of returned complaint devices, product in inventory, and device manufacturing records has identified tool marks isolated to one specific machining mill to be the cause of the fractures. .
Action
Zimmer sent an Urgent Medical Device Recall letter dated November 2014 to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to review the notification and ensure affected personnel were aware of the contents. Customers were advised to quarantine any affected product and notify their Zimmer sales representative who will remove the recalled product from their facility. For patients who previously had this product implanted, it is recommended that they continue their normal operative follow up routine. Customers with questions were instructed to call 1-877-946-2761.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notes supplémentaires dans les données
Intervertebral fusion device with integrated fixation, cervical - Product Code OVE
Cause
A design deficiency was discovered whereby the drill/awl sleeve and spring-loaded drill/awl sleeve instruments (the
sleeves) allow the mating drill and awl instruments to penetrate 1.0 mm beyond their intended depths during screw hole preparation.
Action
The firm, Biomet, sent an "URGENT MEDICAL DEVICE RECALL NOTICE" letter dated 11/11/2014 to their customers. The letter described the product, problem and actions to be taken. The customers were instructed to immediately locate and remove from circulation the items consigned/loaned to your account as identified; carefully follow the instructions on the enclosed "Response Form"; complete and return Response Form via fax to 303.501.8444, and if you have further distributed these items to medical facilities, you MUST notify them of the actions.
Questions related to this notice should be directed to the Sr. Quality Engineer at 518-859-3487 Monday through Friday, 8 am to 4 pm ET.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Two complaints reported that the biolox delta ts ceramic femoral articul/eze12/14 36mm +12 head (product code 1365-36-740) did not fit like the trial; the definite hip construct felt loose. the loosening refers to tissue tension and not the taper connection between the femoral head and stem. the investigations found that the small size sleeve (+1.5 femoral head offset) was incorrectly used in the.
Action
On 11/7/2014, URGENT INFORMATION - RECALL NOTICE notifications were sent to the affected distributors, hospitals, and doctors with instructions for returning the affected product. All distributors were notified via electronic mail, and all hospitals and doctors were notified via sales consultant visit. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. For clinical questions from surgeons, please contact DePuy Orthopaedics, Inc. Scientific Information Office at 1-888-554-2482 (M-F; 8 a.m. - 5 p.m. EDT). For questions about device recall information provided, please contact Kim Earle, Recall Coordinator, at 574-371-4917 (M-F; 8 a.m. - 5 p.m. EDT).
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Complaints have indicated that the attune intuition impaction handle lever has fractured during impaction. if the lever fractures, there is the potential for fractured pieces of the device, or the inner spring, if released, to be left in the patient if not observed during surgery.
Action
On 11/10/2014, URGENT INFORMATION ATTUNE INTUITION IMPACTION HANDLE AND IMPACTORS DEVICE CORRECTION NOTICE notifications were sent to the affected distributors and medical professionals. All distributors were notified via electronic mail, and all medical professionals with affected inventory were also notified by distributor or Sales Consultant via in person delivery or mail or electronic mail. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. For questions, please contact Kim Earle, Recall Coordinator, at 574-371-4917 (M-F; 8 a.m. - 5 p.m. EDT).
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notes supplémentaires dans les données
Bender - Product Code HXW
Cause
Certain lots have the potential for tip breakage while bending 2.0 mm thick plates. if device is used and breaks while in the operating room, surgical delay may occur while alternate bending or cutting instruments are located.
Action
An Urgent Notice: Medical Device Recall, dated December 19, 2014, was sent to end users and sales consultants that described the product, problem, and action to be taken. End users/sales staff were requested to immediately stop use and remove affected product from inventory. Customers were asked to call and obtain a Return Authorization Number, and complete and return the Verification Section of the letter along with the devices to Synthes per the instructions provided. Questions should be directed to 610-719-5450 or your local DePuy Synthes sales consultant.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notes supplémentaires dans les données
Plate, fixation, bone - Product Code HRS
Cause
The minor diameter, major diameter, thread pitch, and pitch on the crosscheck are out of specification, which may result in intraoperative malfunction.
Action
Firm has initiated the recall on 03/06/2015 through written communication via Fed Ex P1 delivery to all Wright Medical Technologies distributors (hubs). Follow-up to non-responding consignees will commence 30 days after initiation. Effectiveness checks will be performed through the confirmation of returned product via a validated ERP system, a returned/signed verification form, and in the case of non-responders, a phone call, e-mail or follow up letter, depending on the region and availability of consignee.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notes supplémentaires dans les données
Plate, fixation, bone - Product Code HRS
Cause
The minor diameter, major diameter, thread pitch, and pitch on the crosscheck are out of specification, which may result in intraoperative malfunction.
Action
Firm has initiated the recall on 03/06/2015 through written communication via Fed Ex P1 delivery to all Wright Medical Technologies distributors (hubs). Follow-up to non-responding consignees will commence 30 days after initiation. Effectiveness checks will be performed through the confirmation of returned product via a validated ERP system, a returned/signed verification form, and in the case of non-responders, a phone call, e-mail or follow up letter, depending on the region and availability of consignee.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notes supplémentaires dans les données
Cerclage, fixation - Product Code JDQ
Cause
The box end label is incorrectly labeled as titanium instead of stainless steel on one lot of implantable songer spinal cable system, single cable, crimp, bar, stainless steel, sterile, distributed by depuy spine inc.
Action
On 5/29/15, RTI Surgical sent DePuy/Synthes Spine Inc. with an URGENT: Medical Device Voluntary Recall Notification identifying the recalled product, reason for recall, potential risks, and immediate actions required.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notes supplémentaires dans les données
Rod, fixation, intramedullary and accessories - Product Code HSB
Cause
The height of one of the three guiding pins of the helical blade inserter for the tfna is oversized. the helical blade inserter is therefore unable to pass through the blade/screw guide sleeve (part 03.037.017).
Action
A recall notification letter, dated June 10, 2015, was sent to sales consultants and Synthes distribution sites requesting return of affected devices.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notes supplémentaires dans les données
Appliance, fixation, spinal intervertebral body - Product Code KWQ
Cause
Shipped to a single consignee prior to completion of final inspection.
Action
Exactech contacted their consignee by telephone on May 21, 2015, of this recall for their two products and instructed him to recover the devices from the hospital. Devices were recovered from the hospital May 22, 2015, and returned to the firm on May 26. All devices have been recovered.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notes supplémentaires dans les données
Appliance, fixation, spinal intervertebral body - Product Code KWQ
Cause
Shipped to a single consignee prior to completion of final inspection.
Action
Exactech contacted their consignee by telephone on May 21, 2015, of this recall for their two products and instructed him to recover the devices from the hospital. Devices were recovered from the hospital May 22, 2015, and returned to the firm on May 26. All devices have been recovered.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notes supplémentaires dans les données
Probe - Product Code HXB
Cause
During cleaning, two separate units of the nexgen tibial base plates did not fit with the tibial base plate handle.
Action
Zimmer, Inc. sent recall letters/ hospital risk managers emails and letters via courier service on December 2, 2014. Inventory Return Certification Forms were also provided. Your responsibilities are:
1. Locate all products from the identified lot above and quarantine them immediately.
2. Carry out a physical count of all affected product in your territory and complete the Inventory Return Certification Form (Attachment 1), and email a completed copy of Attachment 1 to corporatequality.postmarket@zimmer.com.
3. Return the recalled product along with the completed Inventory Return Certification Form (Attachment 1) to the specified address.
4. Please note that any hospitals that received direct shipments of this product from Zimmer will be sent a copy of the Risk Manager letter directly.
5. Please notify Zimmer of any hospitals that you have further distributed the affected product to. Supply the information for any hospitals that you have identified, as well as the affected surgeons using the provided spreadsheet template. Zimmer will notify the additional surgeons and hospitals you identify via Fed Ex mailing. The template will be emailed to you for completion and return to corporatequality.postmarket@zimmer.com.
6. Ensure that a copy of the Risk Managers letter is left with the facilities from which product is removed by providing them with a copy of the letter, if they are not identified as a facility receiving a letter directly from Zimmer.
7. If after reviewing this notification you have further questions or concerns please call the customer call center at 1-877-946-2761 between 8:00 am and 5:00pm EST.
Biomet emailed a new letter with correct distribution dates on 1/16/2015.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Potential for delay in surgery and/or ball bearings to be left in the wound. repeated cycles of sonic cleaning can break down the ball bearing retention feature (swage) on the shim. to prevent this, tasp shims must be manually cleaned and inspected per product insert, 87-6203-991-22.
Action
Zimmer sent an URGENT MEDICAL DEVICE CORRECTION notifications letter dated December 11, 2014 to affected distributors via electronic mail, and all hospital Risk Managers and distributors with affected inventory via courier. The letter identified the affected product, problem, instructions for responding to the formal recall notification.and actions to be taken. For questions call the customer call center at 1-800-348-2759.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notes supplémentaires dans les données
Instrument, surgical, orthopedic, ac-powered motor and accessory/attachment - Product Code HWE
Cause
In may 2013, the original recall was initiated due to complaints the hand switch for the electric pen drive was not labeled correctly, and the on and lock labels were interchanged.
Action
Synthes sent an Urgent Medical Device Recall Notice, dated December 19, 2014, to all affected customers to provide updated information on additional lots identified. Customers were instructed to examine their inventory for the affected product, remove from inventory, and return them to Synthes immediately for replacement. Customers were instructed to call Synthes to obtain a Return Authorization (RA) Number, complete the Verification Section at the end of the letter, and return the Verification Form with the product to: Credit/Returns, Synthes 1101 Synthes Avenue Monument, CO 80132. Customers with questions were instructed to call 610-719-5450 or email FieldAction@synthes.com. For questions regarding this recall call 610-719-5000.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
The recommended sterilization and drying parameters are not effective to achieve the minimum required sterilization level.
Action
FIELD SAFETY NOTICE / PRODUCT NOTIFICATION letters dated December 11, 2014 were sent to all direct accounts (customers). The letters included instructions for customers to utilize the sterilization cycles included in the letter. Customers were further advised to alert users of these devices to the sterilization process changes. Customers with questions can contact Brainlab by telephone at 800-597-5911 or via e-mail at us.support@brainlab.com.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
The taper adaptor is missing from the packaging, which may result in a delay in surgery greater than 30 minutes.
Action
Biomet issued written notices on 1/7/2015, Requested completion and submission of response/effectiveness forms for Product Part Number: 01 0000589 Product Description: Comprehensive Reverse Shoulder Glenosphere Mini
Baseplate 25mm with Taper Adapter, Expiration October 2024. Questions related to this notice should be directed to Audrey Daenzer, Field Action Specialist (574) 372-1570, Monday through Friday, 8 a.m. to 5 p.m.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notes supplémentaires dans les données
Instrument, surgical, orthopedic, ac-powered motor and accessory/attachment - Product Code HWE
Cause
Stryker is recalling system 6 aseptic housing, due to a failed continuous bond line test from a housing without a continual weld. customers are instructed to return recalled product to stryker.
Action
Stryker sent notice to customers via Fed ex overnight on 1/16/2015, regarding its voluntary recall of Stryker System 6 Aseptic Housings because the laser welder may have failed to produce a continuous bond line for the four specific lots: 13205, 13209, 13210, 13212. Customers are instructed to discontinue use, complete the business reply form, and return all recalled product. Customers may contact Kara Spath at 269-389-4518, or kara.spath@stryker.com.