Recalling Organisation: Covidien NZ Ltd, Level 3, Building 5, Central Park Corporate Centre, 666 Great South Road, Penrose, AUCKLAND 1051
Cause
The manufacturer has received customer reports of ht70 series ventilators alarming and going to internal backup battery sooner than expected while the ventilator is being operated on the power pac battery.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Notes supplémentaires dans les données
Recalling Organisation: Siemens Healthcare (NZ) Ltd, Millennium Centre, Part Level 2, Building A, 600 Great South Road, Ellerslie, AUCKLAND 1051
Cause
Siemens has determined that the rheumatoid factor ssvs currently assigned to this lot may cause an average -15% shift in quality control and sample recovery.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Notes supplémentaires dans les données
Recalling Organisation: Johnson & Johnson Medical New Zealand Ltd, 507 Mt Wellington Highway, Mt Wellington, AUCKLAND 1060
Cause
The manufacturer has received reports of smoke emanating from the system and upon further investigation has discovered that a circuit board malfunction may occur in the channel control manifold sub-assembly.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Notes supplémentaires dans les données
Recalling Organisation: Johnson & Johnson Medical New Zealand Ltd, 507 Mt Wellington Highway, Mt Wellington, AUCKLAND 1060
Cause
The manufacturer has recently determined that under certain circumstances, the evotech endoscope cleaner and reprocessor system may not properly calculate the expiration date of the cidezyme xtra multi enzymatic detergent from the information entered by the user. in these circumstances, the cidezyme xtra multi enzymatic detergent expiration date displayed on the evotech screen is one month after the labeled expiration date.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Synthes has two couplings in place for dhs/dcs® screws: standard with teethed recess and octagonal recess. the pictogram on the labels of the specified devices shows the wrong coupling. all the other information on the label is correct.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Synthes has two couplings in place for dhs/dcs® screws: standard with teethed recess and octagonal recess. the pictogram on the labels of the specified devices shows the wrong coupling. all the other information on the label is correct.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Notes supplémentaires dans les données
Recalling Organisation: Johnson & Johnson Medical New Zealand Ltd, 507 Mt Wellington Highway, Mt Wellington, AUCKLAND 1060
Cause
During post market surveillance of the enseal g2 superjaw the team observed variability in how surgeons were using the device, which led to clarification of the ifu.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
ICU Medical, Inc. acquired from Pfizer in early 2017, Hospira Infusion Systems, the portion of Hospira dedicated to develop infusion pumps.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Latest internal stability data for seraclone-anti-s showed a reactivity decrease of the lot numbers listed. weak or false negative results might be possible in testing., affected batch had only been dstributed to one customer in new zealand - nzbs.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
ICU Medical, Inc. acquired from Pfizer in early 2017, Hospira Infusion Systems, the portion of Hospira dedicated to develop infusion pumps.
Failure to close the slide clamp when the syringe was maipulated or when stopping infusion during the delivery of fluids. this may lead to a potental over-delivery of medication.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Notes supplémentaires dans les données
Recalling Organisation: GE Healthcare Limited, 8 Tangihua Street, Auckland 1010
Cause
There is a potential safety issue that may occur when direct multi-planar reformat (dmpr) series and screen saved images are sent to some pacs. when sent to pacs there may be a problem with how these types of series are uniquely identified on the pacs.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
During beam delivery, the tolerance values are used to do a check for unwanted movements. these values are known as "static tolerances" which are related to all the digital accelerator geometric movement axes except table column. the problem is that the "static tolerances" from the calibration files (loaded database) are used for beam delivery instead of the machine calculated values. a fault occurs if the loaded database is incorrect.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
The manufacturer has received reports of false occlusion alarms, resulting in the termination of infusion runs for these products when they are used to deliver medication via an infusion pump or syringe driver.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Some icds that have previously exhibited a false battery depletion alert will display another false battery depletion alert despite having received the software upgrade to address this issue.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Following a review of clinical data from the manufacturer's 5-year global study the instructions for use for this device are being updated with new guidance for patient selection, and patient monitoring.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Notes supplémentaires dans les données
Recalling Organisation: Obex Medical Ltd, 303 Manukau Road, Epsom, AUCKLAND
Cause
Cook medical has received a number of complaints relating to the delivery system for the cook zilver ptx drug eluting stent involving fractures of the delivery sytem inner catheter after stent deployment, and separation of the inner catheter tip section.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
The product description on the six languages label on the back of the box is wrong in spanish and portuguese. instead of right, the plate is described as being a left plate in these two languages.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Notes supplémentaires dans les données
Recalling Organisation: Siemens Healthcare (NZ) Ltd, Millennium Centre, Part Level 2, Building A, 600 Great South Road, Ellerslie, AUCKLAND 1051
Cause
The manufacturer advises that the laterality information in mammography images is not displayed and these images may not be properly aligned to mammography segments.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Notes supplémentaires dans les données
Recalling Organisation: Philips New Zealand Commercial Limited, Level 3, 123 Carlton Gore Road, Newmarket, AUCKLAND 1023
Cause
The philips heartstart mrx monitor/defibrillator may fail to deliver defibrillation therapy in either manual defib or aed mode. if this occurs, the mrx may simultaneously display a "no shock delivered" message along with a "shock equip malfunction" inop and a red x in the rfu indicator. in addition, the pads ecg waveform may display a non-physiologic flat line rhythm.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Specific lots of some collagen products may have deviated from a production process. the affected lots passed all product finished goods testing, are sterile, and were tested and accepted for endotoxin levels. however, due to a process deviation, they may have been released with higher levels of endotoxins that permitted by release specifications. higher endotoxin levels may result in fever in the immediate postoperative period. the manufacturer are not aware of any reports of patient injuries or other adverse events in connection with these product lots., the health hazard evaluation (hhe) provided by the manufacturer details that during an internal quality assurance review of processes, specific lots manufactured during the periods december 2010 to may 2011, and then december 2012 to february 2013, may possibly be affected.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Specific lots of some collagen products may have deviated from a production process. the affected lots passed all product finished goods testing, are sterile, and were tested and accepted for endotoxin levels. however, due to a process deviation, they may have been released with higher levels of endotoxins that permitted by release specifications. higher endotoxin levels may result in fever in the immediate postoperative period. the manufacturer are not aware of any reports of patient injuries or other adverse events in connection with these product lots., the health hazard evaluation (hhe) provided by the manufacturer details that during an internal quality assurance review of processes, specific lots manufactured during the periods december 2010 to may 2011, and then december 2012 to february 2013, may possibly be affected.