Recalling Organisation: Johnson & Johnson (NZ) Limited, 13a Gabador Place, Mt Wellington, AUCKLAND
Cause
It was reported that the drive shaft, tube assembly, and reamer head have the potential to break when inccorrectly assembled or used improperly. precautionary statements are being added to the "reamer / irrigator / aspirator (ria) surgical technique guide".
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Notes supplémentaires dans les données
Recalling Organisation: Johnson & Johnson (NZ) Limited, 13a Gabador Place, Mt Wellington, AUCKLAND
Cause
It was reported that the drive shaft, tube assembly, and reamer head have the potential to break when inccorrectly assembled or used improperly. precautionary statements are being added to the "reamer / irrigator / aspirator (ria) surgical technique guide".
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Notes supplémentaires dans les données
Recalling Organisation: REM Systems Limited, 69 Elizabeth Knox Place, Glen Innes, AUCKLAND
Cause
The targeting jig mast is separating from the jig boom. any separation will impact alignment and can affect the ability to fixate the screws.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
During the manufacture of the architect b12 reagent kits (list 7k61) listed above, an incorrect ifu (instructions for use) commodity number g3-0354/r04 may have been inserted into the reagent kits.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
The manufacturer has received complaints from customers that when using the epidural mode with patient bolus, the amount of medication delivered may be greater than programmed. q core medical is adding to the user manual the parameters of the reference catheter to allow precise correlation between the catheter dimensions, the back pressure and the deviation in flow that the catheter may cause.
Peruvian data is current through August 2018. All of the data comes from Digemid, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Peru.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Notes supplémentaires dans les données
Recalling Organisation: Siemens Healthcare (NZ) Ltd, Millennium Centre, Part Level 2, Building A, 600 Great South Road, Ellerslie, AUCKLAND 1051
Cause
A possible position sensor fault in the swivel base axis, not detected by the system software, may cause the c-arm system to move faster than normal when movement is enabled by the operator. the system will not move of its own accord. movement must still be initiated by the operator, however, the system can unexpectedly exceed its usual speed, potentially resulting in a collision.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
The manufacturer has identified that when the operator detaches the fixation ring from the paper liner, in some cases, fractions of the paper liner's surface coating sticks to the adhesive surface of the ring.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Notes supplémentaires dans les données
Recalling Organisation: Obex Medical Ltd, 303 Manukau Road, Epsom, AUCKLAND
Cause
Cook medical has received reports of catheter tip splits and or separation., oct-2015: cook extended the recall to cover additional catheter models and all lots in current distribution.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Notes supplémentaires dans les données
Recalling Organisation: Obex Medical Ltd, 303 Manukau Road, Epsom, AUCKLAND
Cause
Cook medical has received reports of catheter tip splits and or separation., oct-2015: cook extended the recall to cover additional catheter models and all lots in current distribution.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Notes supplémentaires dans les données
Recalling Organisation: Obex Medical Ltd, 303 Manukau Road, Epsom, AUCKLAND
Cause
Cook medical has received reports of catheter tip splits and or separation., oct-2015: cook extended the recall to cover additional catheter models and all lots in current distribution.
Peruvian data is current through August 2018. All of the data comes from Digemid, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Peru.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Notes supplémentaires dans les données
Recalling Organisation: Siemens Healthcare (NZ) Ltd, Millennium Centre, Part Level 2, Building A, 600 Great South Road, Ellerslie, AUCKLAND 1051
Cause
The manufacturer has been advised of a report of unintended detector motion caused by the simultaneous failure of two components. this unique combination of failures caused the axis brake to be ineffective, leading to the detector motion. siemens will install an improved brake control circuit to correct this condition.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Notes supplémentaires dans les données
Recalling Organisation: Obex Medical Ltd, 303 Manukau Road, Epsom, AUCKLAND
Cause
The manufacturer has determined that validation data related to cleaning instructions and sterilization methods and the methods defined in instructions for use (ifu) for the sureflex™ reusable fibers and the reusable stripper and cleaver accessories do not meet the new guidance on qualifying reusable products.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Notes supplémentaires dans les données
Recalling Organisation: Obex Medical Ltd, 303 Manukau Road, Epsom, AUCKLAND
Cause
The manufacturer has determined that validation data related to cleaning instructions and sterilization methods and the methods defined in instructions for use (ifu) for the sureflex™ reusable fibers and the reusable stripper and cleaver accessories do not meet the new guidance on qualifying reusable products.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Notes supplémentaires dans les données
Recalling Organisation: Obex Medical Ltd, 303 Manukau Road, Epsom, AUCKLAND
Cause
The manufacturer has determined that validation data related to cleaning instructions and sterilization methods and the methods defined in instructions for use (ifu) for the sureflex™ reusable fibers and the reusable stripper and cleaver accessories do not meet the new guidance on qualifying reusable products.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Notes supplémentaires dans les données
Recalling Organisation: Beckman Coulter NZ Ltd, Suite 2, Level 4, 30 St Benedicts Street, Newton, AUCKLAND
Cause
Iris has confirmed the ichem velocity automated urine chemistry systems do not maintain the on-board strip stability claim of 10 days at 18-28 degrees c at 20-80% relative humidity as stated in the product labeling. the on-board strip stability is affected by temperature and humidity in the laboratory. if a laboratory operates outside of the defined parameters , test results may be affected as indicated below:, * false low or false negative results: ketone, blood, * false high or false positive results: bilirubin, urobilinogen, glucose, nitrite, * false high results: leukocytes, ascorbic acid, protein and ph are not affected.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Notes supplémentaires dans les données
Recalling Organisation: Medtronic New Zealand Limited, Unit 16 Mezzanine Level, 5 Gloucester Park Road, Onehunga, AUCKLAND
Cause
Following monitoring of reported events, clinical trials and published literature, medtronic has updated the warnings and adverse events sections. the labelling updates potential risks which have been reported with deep brain stimulation therapy.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Notes supplémentaires dans les données
Recalling Organisation: Medtronic New Zealand Limited, Unit 16 Mezzanine Level, 5 Gloucester Park Road, Onehunga, AUCKLAND
Cause
Following monitoring of reported events, clinical trials and published literature, medtronic has updated the warnings and adverse events sections. the labelling updates potential risks which have been reported with deep brain stimulation therapy.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Notes supplémentaires dans les données
Recalling Organisation: Medtronic New Zealand Limited, Unit 16 Mezzanine Level, 5 Gloucester Park Road, Onehunga, AUCKLAND
Cause
Following monitoring of reported events, clinical trials and published literature, medtronic has updated the warnings and adverse events sections. the labelling updates potential risks which have been reported with deep brain stimulation therapy.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Notes supplémentaires dans les données
Recalling Organisation: Medtronic New Zealand Limited, Unit 16 Mezzanine Level, 5 Gloucester Park Road, Onehunga, AUCKLAND
Cause
Following monitoring of reported events, clinical trials and published literature, medtronic has updated the warnings and adverse events sections. the labelling updates potential risks which have been reported with deep brain stimulation therapy.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Notes supplémentaires dans les données
Recalling Organisation: Medtronic New Zealand Limited, Unit 16 Mezzanine Level, 5 Gloucester Park Road, Onehunga, AUCKLAND
Cause
Following monitoring of reported events, clinical trials and published literature, medtronic has updated the warnings and adverse events sections. the labelling updates potential risks which have been reported with deep brain stimulation therapy.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Notes supplémentaires dans les données
Recalling Organisation: Medtronic New Zealand Limited, Unit 16 Mezzanine Level, 5 Gloucester Park Road, Onehunga, AUCKLAND
Cause
Following monitoring of reported events, clinical trials and published literature, medtronic has updated the warnings and adverse events sections. the labelling updates potential risks which have been reported with deep brain stimulation therapy.