Cause
One complaint regarding each issue has been received. complaint states the dpb1 ssp unitray kit has a false negative in lane 30 which gives no perfect match typing result for the dpb1*29:01 allele. internal investigation has confirmed with the now available rare sample type that lane 30 is negative for the dpb1 *29 :01 allele in the listed affected kits according to the labeling provided with the kits. the primer mixes had not been tested with a dpb1 *29:01 sample the labeling stated the primers in lane 30 would amplify the targeted sequences in the dpb1 *29:01 allele.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notes supplémentaires dans les données
Cause
Stryker instruments has revised the instructions for use (ifu) for the painpump2 block-aid.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notes supplémentaires dans les données
Cause
Siemens has confirmed a report of leaking flex reagent cartridges from tpsa lot 113300bb. if a leaking flex reagent cartridge is used test may be falsely depressed. the magnitude of the inaccuracy and the number of test results affected may vary based on the extent of leaking.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notes supplémentaires dans les données
Cause
There is the possibility that a clip in the applier may be missing a boss. a boss is the part of the clip that is used to hold the clip within the applier jaws prior to application.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notes supplémentaires dans les données
Cause
Bard is notifying customers of incorrect lumen gauge size labeling on some per-q-cath products. bard identifies catheter size in two ways on this product by french size and gauge size. the french size is not affected and is correct for these products. product is not required to be removed as the product functions as intended and is manufactured according to design specifications with the correct french size designation.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notes supplémentaires dans les données
Cause
Argon has received a complaint from a foreign distributor of a hole in the blister package. could result in loss of product sterility and potential for patient infection during product use.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notes supplémentaires dans les données
Cause
Impac determined that a scenario existed whereby certain records could be cross-linked. the problem was diagnosed and verfied as an occurance through testing at the customer site.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notes supplémentaires dans les données
Cause
Calaxo screws may present a post-operative condition "pretibial soft tissues swelling" after implantation including the potential for graft failure and premature material degradation.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notes supplémentaires dans les données
Cause
Leg bolts may be longer than required which can result in fracture and loosening of bolts.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notes supplémentaires dans les données
Cause
Size 10 lexicon k-files (kf00825) were packed and labelled as size 8. the markings on the actual files correctly indicate a size 10.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notes supplémentaires dans les données
Cause
An inconsistency was noted in the i-stat system manuals regarding the volume of saline used to flush an indwelling line and the volume of blood to be discarded for clearing the line when testing coagulation cartridges.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notes supplémentaires dans les données
Cause
Performance does not meet sample stability claims. samples stored at room temperature and run 24 hr after collection may yield erroneously high rdw and m0% and erroneously low gr%.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notes supplémentaires dans les données
Cause
Calibration curve failure and high recovery for biorad dade and msa controls at low control levels.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notes supplémentaires dans les données
Cause
Software update vb20a to vb20b. this correction is a performance issue not related to patient or operator safety. the deadline to complete this modification is february 26 2003.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notes supplémentaires dans les données
Cause
Some eclipse oxygen system (model 1000) units may experience premature wear on an electrical connector between the battery bridge printed circuit board and the power manager printed circuit board that could lead to a temporary power interruption when operating from the power cartridge.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notes supplémentaires dans les données
Cause
Manufacturer has documented a potential risk of data loss when affected base units (used in parallel with point of care instrumentation) are using a usb connection for data transmission to a data management system (cobas it1000 or any other 3rd party dms). although the occurrence is rare in the worst case scenario this issue could lead to an erroneous assignment of patient data.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notes supplémentaires dans les données
Cause
The error code displayed on the screen can lead to misinterpretation which could potentially delay the treatement of patient with very low blood glucose.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notes supplémentaires dans les données
Cause
Discrepacy on workin probe suspension an substrate stability information in package inserts/method manuals. (9 working substrates affected.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notes supplémentaires dans les données
Cause
Punches were not manufactured to our specifications and the tip of the punch is welded on and not one piece as specified. this can cause the tip of the punch to break off during use.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notes supplémentaires dans les données
Cause
System has the potential for an internal component to overheat which could pose the risk of smoke and fire.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notes supplémentaires dans les données
Cause
The aformentioned lots will not support the growth of haemophilus parahaemolyticus when used in conjunction with x+v taxo factor strips.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notes supplémentaires dans les données
Cause
Possible angulation problem which may cause the bending rubber to partially separate over the bending section of the scope which may cause exposure of the metal section beneath.