US: Nationwide OUS: Canada, Afghanistan Albania Algeria Andorra Argentina Australia Austria Azerbaijan Bahamas Bangladesh Belarus Belgium Bolivia Bosnia and Herzegovina Botswana Brazil Burkina Faso Chile China Colombia Columbia Congo, Democratic Costa Rica Croatia Cyprus Czech Republic Denmark Dominican Republic Ecuador Egypt Estonia Ethiopia Finland France Georgia Germany Ghana Greece Guadeloupe Hong Kong Hungary Iceland India Indonesia Iran Iraq Ireland Israel Italy Japan Jersey Jordan Kazakhstan Kenya Korea, Republic of Kuwait Kyrgyzstan Latvia Lebanon Libya Lithuania Luxembourg Malaysia Malta Martinique Mauritius Mexico Monaco Mongolia Morocco Mozambique Nepal Netherlands New Caledonia New Zealand Norway Oman Pakistan Palestinian territory, occupied Panama Papua New Guinea Peru Philippines Poland Portugal Puerto Rico Romania Russian Federation Rwanda Saudi Arabia Serbia Singapore Slovakia Slovenia South Africa Spain Sri Lanka Sudan Sweden Switzerland Syrian Arab Republic Taiwan Tanzania Thailand Tunisia Turkey Uganda Ukraine United Arab Emirates United Kingdom Venezuela Viet Nam Yemen Zimbabwe
Description du dispositif
SENSE Body Coil 1.5T with identification; used in the ACS NT, Intera 1.5T and Achieva 1.5T MR system
MR systems have a strong magnetic field around the magnet, the so called fringe field. Reduction of this fringe field (e.g. 0.5-mT area) can be required under certain circumstances. A reduction can be achieved with ferromagnetic plates mounted to the parent walls, under the construction floor and above the RF-ceiling. These plates are called passive shielding.
Passive shielding calculations require dedicated Philips Healthcare software. As documented in the Planning Reference Documentation, these calculations must be performed by Philips. It provides the requirements for the thickness and the location of the shielding material which are to be implemented, preferably by Philips’ selected RF cage suppliers.