Rappel de 5mm Round Fluted Bur Aggressive, Round Fluted Soft Touch 3.0mm, Round Fluted Bur 6.0mm, 3mm Round Fluted Bur Aggressive, Neuro Drill 3.0mm, 6mm Round Fluted Bur Aggressive, Round Fluted Bur 1.0mm and 5.0mm Egg Bur (Orthopaedic bur used for soft tissue debridement and bone abrasion and removal)

Selon Department of Health, Therapeutic Goods Administration, ce/cet/cette rappel concerne un dispositif en/au/aux/à Australia qui a été fabriqué par Stryker Australia Pty Ltd.

Qu'est-ce que c'est?

Une correction ou un retrait opéré par un fabricant afin de répondre à un problème causé par un dispositif médical. Les rappels surviennent lorsqu'un dispositif médical est défectueux, lorsqu'il pourrait poser un risque pour la santé, ou les deux à la fois.

En savoir plus sur les données ici
  • Type d'événement
    Recall
  • ID de l'événement
    RC-2014-RN-00593-1
  • Classe de risque de l'événement
    Class II
  • Date de mise en oeuvre de l'événement
    2014-05-30
  • Pays de l'événement
  • Source de l'événement
    DHTGA
  • URL de la source de l'événement
  • Notes / Alertes
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Notes supplémentaires dans les données
  • Cause
    During manufacture, the notching completed on one specific machine may have been started too close to the end of the shank. in some products (approximately 1/1000), there may be missing notches, partial notches or no notches at all. the listed cutting accessories are manufactured with notches at the proximal end of the bur. the notch is a feature on the bur that is used to retain and control the exposure of the bur within the drill/attachment. the correct number of notches varies from 2 to 5, based on the specific product number.
  • Action
    Stryker is requesting their customers to quarantine all affected stock and advising that they will be contacted to organize for the return of the product.

Device

  • Modèle / numéro de série
    5mm Round Fluted Bur Aggressive, Round Fluted Soft Touch 3.0mm, Round Fluted Bur 6.0mm, 3mm Round Fluted Bur Aggressive, Neuro Drill 3.0mm, 6mm Round Fluted Bur Aggressive, Round Fluted Bur 1.0mm and 5.0mm Egg Bur (Orthopaedic bur used for soft tissue debridement and bone abrasion and removal)Multiple item numbers and lot numbers affected
  • Manufacturer

Manufacturer

  • Société-mère du fabricant (2017)
  • Source
    DHTGA