Rappel de Active Knee System - Patella Insert (Used as a component in total knee replacement) ASDM

Selon Department of Health, Therapeutic Goods Administration, ce/cet/cette rappel concerne un dispositif en/au/aux/à Australia qui a été fabriqué par Advanced Surgical Design & Manufacture Limited.

Qu'est-ce que c'est?

Une correction ou un retrait opéré par un fabricant afin de répondre à un problème causé par un dispositif médical. Les rappels surviennent lorsqu'un dispositif médical est défectueux, lorsqu'il pourrait poser un risque pour la santé, ou les deux à la fois.

En savoir plus sur les données ici
  • Type d'événement
    Recall
  • ID de l'événement
    RC-2012-RN-01164-1
  • Classe de risque de l'événement
    Class II
  • Date de mise en oeuvre de l'événement
    2012-11-27
  • Pays de l'événement
  • Source de l'événement
    DHTGA
  • URL de la source de l'événement
  • Notes / Alertes
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Notes supplémentaires dans les données
  • Cause
    Advanced surgical design and manufacture has become aware of elevated failure rates of the patella component of the active knee system, specifically related to the shearing of the pegs. fourteen occurrences have been reported to date, which represents a failure rate of 0.7% of patellae implanted since july 2007. batches which have been found to be affected were manufactured between 2007 and 2011, with failures presenting between 1 to 4 years after implantation.Electron microscopy of the failed patella has shown that the patella pegs are failing due to fatigue. the initiation site for the fatigue failure appears to be a small overhang artefact, between 10 to 30 microns in depth, at the base of the patella peg. this small artefact may lead to elevated stresses at the base of the peg, and therefore increases the risk of premature product failure.
  • Action
    Advanced Surgical Design and Manufacture is advising implanting/treating surgeons on how to manage patients implanted with the affected Active Knee System. For more details, please see http://www.tga.gov.au/safety/alerts-device-active-knee-121205.htm This action has been closed-out on 12/04/2016.

Device

  • Modèle / numéro de série
    Active Knee System - Patella Insert (Used as a component in total knee replacement) ASDM Part Number: 10-1514-075XLots manufactured prior to June 2012ARTG Number:133900
  • Manufacturer

Manufacturer