Rappel de ADT1018 and ADT1018-50 Flow QC Tubing – haemodialysis blood tubing Set

Selon Department of Health, Therapeutic Goods Administration, ce/cet/cette rappel concerne un dispositif en/au/aux/à Australia qui a été fabriqué par Emergo Asia Pacific Pty Ltd T/a Emergo Australia.

Qu'est-ce que c'est?

Une correction ou un retrait opéré par un fabricant afin de répondre à un problème causé par un dispositif médical. Les rappels surviennent lorsqu'un dispositif médical est défectueux, lorsqu'il pourrait poser un risque pour la santé, ou les deux à la fois.

En savoir plus sur les données ici
  • Type d'événement
    Recall
  • ID de l'événement
    RC-2018-RN-00220-1
  • Classe de risque de l'événement
    Class I
  • Date de mise en oeuvre de l'événement
    2018-03-13
  • Pays de l'événement
  • Source de l'événement
    DHTGA
  • URL de la source de l'événement
  • Notes / Alertes
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Notes supplémentaires dans les données
  • Cause
    Emergo has been notified by the manufacturer, transonic systems (transonic), that the sterility of adt1018 flow qc tubing set cannot be assured. transonic is not aware of any actual products that are not sterilized; however, there is a small risk that unsterilized product may have been released. given this possibility, in the event the adt1018 flow qc tubing were to be used without being sterilized, this would increase the risk of infection post use for patients who have come into contact with any unsterilized product. to date, no complaints or adverse events have been reported to transonic related to this issue.
  • Action
    Transonic has discontinued the ADT1018 product line and has ceased purchasing product from this supplier entirely. The ADT1010 tubing set replaces ADT1018 Flow QC tubing. Emergo is advising users to immediately discontinue use and distribution of the affected product. Return or destroy the affected product and document that destruction on the Medical Device Recall Return Response Form provided with the Customer Letter. Upon receipt of the completed form a product replacement will be issued. If it appears that any patients have had contact with the ADT1018 Flow QC tubing set, please observe these patients for any signs of infection as per your hospital protocol and report if any such events take place.

Device

  • Modèle / numéro de série
    ADT1018 and ADT1018-50 Flow QC Tubing – haemodialysis blood tubing Set Catalogue Number: ADT1018Lot Numbers: B151130E2 and B161115E0Expiry: 30/11/2018 and 15/11/2019Catalogue Number: ADT1018-40Lot Numbers: B151130E2 and B161115E0Expiry: 30/11/2018 and 15/11/2019ARTG Number 187558
  • Classification du dispositif
  • Manufacturer

Manufacturer