Rappel de ADVIA Centaur and ADVIA Centaur XP (An in vitro diagnostic medical device (IVD))

Selon Department of Health, Therapeutic Goods Administration, ce/cet/cette rappel concerne un dispositif en/au/aux/à Australia qui a été fabriqué par Siemens Ltd Diagnostics Division.

Qu'est-ce que c'est?

Une correction ou un retrait opéré par un fabricant afin de répondre à un problème causé par un dispositif médical. Les rappels surviennent lorsqu'un dispositif médical est défectueux, lorsqu'il pourrait poser un risque pour la santé, ou les deux à la fois.

En savoir plus sur les données ici
  • Type d'événement
    Recall
  • ID de l'événement
    RC-2014-RN-01116-1
  • Classe de risque de l'événement
    Class II
  • Date de mise en oeuvre de l'événement
    2014-10-15
  • Pays de l'événement
  • Source de l'événement
    DHTGA
  • URL de la source de l'événement
  • Notes / Alertes
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Notes supplémentaires dans les données
  • Cause
    Siemens healthcare diagnostics has identified an issue that on some systems a slight cracking of the tubing (the tubing for the wash 1, acid, and base bottle reservoirs) at the connecting fitting on the reservoir connector assembly may occur due to mechanical stress. the cracking may result in a minor drip of fluid onto the reservoir and into the drip tray located underneath the reservoirs. the potential for minimal fluid leakage outside the system exists. the drip will cause the reservoir to empty faster than the system expects. once the acid and base fluids are depleted, the system detects it and will flag all subsequent results with a signal error flag, once the lack of acid/base solution is detected. when wash 1 fluid becomes low, the system will detect it and stop processing samples. in this case, a limited number of test results (up to 20 tests) processed prior to a signal error may be affected and not flagged.
  • Action
    Siemens is advising the customers to examine their system to see if there is any fluid in the reservoir drip tray. If fluid is present, the customer is advised to contact their Siemens Customer Care Center or local Siemens technical support representative to schedule a visit. Until a service engineer has inspected their system for this issue, Siemens recommends that the customers replace the bulk bottle for all fluids immediately when the yellow warning appears. This will help mitigate the possibility of improper fluid dispenses. Siemens service engineers will inspect the tubing on the next service visit and replace the tubing if necessary. This action has been closed-out on 11/08/2016.