Rappel de Aespire 7900, Aespire View, Aestiva 7900, Aestiva MRI, Aisys, Aisys CS2, Avance, and Avance CS2 Anesthesia devices installed from April 20, 2015 through October 2015. Service kits containing the ventilator Drive Gas Check Valve shipped from April 20, 2015 through October 2015 are also affected

Selon Department of Health, Therapeutic Goods Administration, ce/cet/cette rappel concerne un dispositif en/au/aux/à Australia qui a été fabriqué par GE Healthcare Australia Pty Ltd.

Qu'est-ce que c'est?

Une correction ou un retrait opéré par un fabricant afin de répondre à un problème causé par un dispositif médical. Les rappels surviennent lorsqu'un dispositif médical est défectueux, lorsqu'il pourrait poser un risque pour la santé, ou les deux à la fois.

En savoir plus sur les données ici
  • Type d'événement
    Recall
  • ID de l'événement
    RC-2016-RN-00038-1
  • Classe de risque de l'événement
    Class II
  • Date de mise en oeuvre de l'événement
    2016-01-12
  • Pays de l'événement
  • Source de l'événement
    DHTGA
  • URL de la source de l'événement
  • Notes / Alertes
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Notes supplémentaires dans les données
  • Cause
    The drive gas check valve of the affected devices could become stuck in a fixed open position which could cause pressure to build in the mechanical ventilation cycle. if this issue is left unresolved, it could result in excessive or prolonged pressure in the patient breathing circuit during ventilation potentially resulting in barotrauma. there have been no injuries reported as a result of this issue.
  • Action
    The sponsor is advising users that the affected devices can continue to be used. In the event the reported issue does occur, the affected device provides alarms and there are automatic inherent safety mitigations by design to help ensure patient safety. The sponsor will contact users to arrange for an on-site correction. This action has been closed-out on 03/02/2017.

Device

  • Modèle / numéro de série
    Aespire 7900, Aespire View, Aestiva 7900, Aestiva MRI, Aisys, Aisys CS2, Avance, and Avance CS2 Anesthesia devices installed from April 20, 2015 through October 2015. Service kits containing the ventilator Drive Gas Check Valve shipped from April 20, 2015 through October 2015 are also affectedService kit part numbers: 1009-8216-000, 1503-3006-000, 1503-8102-000, 1009-8423-000 and 1503-8101-000ARTG Number: 93955