Rappel de Airsep Impulse Elite Oxygen Conserving Device

Selon Department of Health, Therapeutic Goods Administration, ce/cet/cette rappel concerne un dispositif en/au/aux/à Australia qui a été fabriqué par Air Liquide Healthcare Pty Ltd.

Qu'est-ce que c'est?

Une correction ou un retrait opéré par un fabricant afin de répondre à un problème causé par un dispositif médical. Les rappels surviennent lorsqu'un dispositif médical est défectueux, lorsqu'il pourrait poser un risque pour la santé, ou les deux à la fois.

En savoir plus sur les données ici
  • Type d'événement
    Recall
  • ID de l'événement
    RC-2017-RN-01448-1
  • Classe de risque de l'événement
    Class II
  • Date de mise en oeuvre de l'événement
    2017-12-05
  • Pays de l'événement
  • Source de l'événement
    DHTGA
  • URL de la source de l'événement
  • Notes / Alertes
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Notes supplémentaires dans les données
  • Cause
    An oxygen conserving device (ocd) is subject to sudden and unpredictable failure when connected to a medical oxygen gas cylinder and pressurised. the cause of failure is due to fatigue and aging of the device as the device is now being operated outside the manufacturer's maximum recommended service life of 7yrs. failure consists of sudden localised pressure loss via the pressure gauge leading to mechanical damage. there is a small chance that this mode of failure could result in the device's outer casing breaking apart to a degree where small plastic projectiles could imact the patient depending on their proximity to the device at the time of failure. this make and model ocd (see above) was discontinued by the manufacturer (airsep) in late 2009 and no new units have been supplied to the market since that time.
  • Action
    If you own an AirSep Impulse Elite OCD complete and return the supplied Facsimile Reply Form to arrange for the renewal of your OCD. Return the form even if you no longer have an AirSep Impulse Elite OCD or previously had one but are no longer in possession of it. If you have provided any units to other users please pass this letter onto them and inform Air Liquide. If you rent an AirSep Impulse Elite OCD from AIR LIQUIDE Healthcare, we will be contacting you shortly to arrange a suitable time to recover and replace this device with a newer make and model OCD. Affected units may continue to be used until swapped out subject to the precautions described in the letter to customers.

Device

  • Modèle / numéro de série
    Airsep Impulse Elite Oxygen Conserving DeviceModel Number: OC006-3ARTG Number: 81737
  • Manufacturer

Manufacturer