Rappel de Alaris Infusion pump administration sets

Selon Department of Health, Therapeutic Goods Administration, ce/cet/cette rappel concerne un dispositif en/au/aux/à Australia qui a été fabriqué par CareFusion Australia 316 Pty Ltd.

Qu'est-ce que c'est?

Une correction ou un retrait opéré par un fabricant afin de répondre à un problème causé par un dispositif médical. Les rappels surviennent lorsqu'un dispositif médical est défectueux, lorsqu'il pourrait poser un risque pour la santé, ou les deux à la fois.

En savoir plus sur les données ici
  • Type d'événement
    Recall
  • ID de l'événement
    RC-2018-RN-00722-1
  • Classe de risque de l'événement
    Class II
  • Date de mise en oeuvre de l'événement
    2018-06-19
  • Pays de l'événement
  • Source de l'événement
    DHTGA
  • URL de la source de l'événement
  • Notes / Alertes
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Notes supplémentaires dans les données
  • Cause
    The bd alaris infusion pump administration sets require priming prior to use as described in the directions for use (dfu). bd has identified a potential issue with the bd alaris infusion pump administration sets manufactured during january 2017 through march 2018. the issue pertains to the back check valve and the ability to prime the administration sets. the manufacturer has determined the root causes leading to this issue are associated with: 1. excessive solvent used to connect the tubing to the back check valve leading to an obstruction of the fluid path, thus preventing the priming of the set. 2. a higher “cracking” pressure of the back check valve.To date, no injuries have been reported as a result of this issue.
  • Action
    CareFusion is advising that users can continue to use the product. When a set cannot be appropriately primed, users may be able to increase the pressure by gently squeezing the IV bag following these steps: 1. If no drops are observed, gently squeeze the bag at the fluid level with one hand once. Observe if set is able to prime. 2. If no drops are observed repeat this up to 3 times. 3. If this method is not effective discard and replace the administration set. 4. Consider having additional administration sets available in clinical areas where high risk infusions are administered.

Device

  • Modèle / numéro de série
    Alaris Infusion pump administration sets Product Codes: 2120-0500 , 2420-0007 , 2420-0500, 10015489, 11426965 ARTG Number:125916
  • Manufacturer

Manufacturer