Rappel de BrightView XCT(SPECT/CT diagnostic imaging system)

Selon Department of Health, Therapeutic Goods Administration, ce/cet/cette rappel concerne un dispositif en/au/aux/à Australia qui a été fabriqué par Philips Electronics Australia Ltd.

Qu'est-ce que c'est?

Une correction ou un retrait opéré par un fabricant afin de répondre à un problème causé par un dispositif médical. Les rappels surviennent lorsqu'un dispositif médical est défectueux, lorsqu'il pourrait poser un risque pour la santé, ou les deux à la fois.

En savoir plus sur les données ici
  • Type d'événement
    Recall
  • ID de l'événement
    RC-2013-RN-00944-1
  • Classe de risque de l'événement
    Class I
  • Date de mise en oeuvre de l'événement
    2013-09-06
  • Pays de l'événement
  • Source de l'événement
    DHTGA
  • URL de la source de l'événement
  • Notes / Alertes
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Notes supplémentaires dans les données
  • Cause
    Philips received a report from the field that the flat panel detector (fpd) sensor reported to the operator that the panel was unlatched at the 0 degree position during setup. the operator attempted to manually relatch the fpd as it remained in the 0 degree position. in trying to restow the fpd, it fell from its stowed position and contacted the patient's leg and the technologist's arm.
  • Action
    Philips Healthcare is advising their customers to only deploy or stow the FPD when in the +90 degrees position. Philips will be implementing a field correction to prevent the issue from occurring.