Rappel de C-series carry bars (used with C300, C450 Manual, C625 ceiling lifts/hoists)

Selon Department of Health, Therapeutic Goods Administration, ce/cet/cette rappel concerne un dispositif en/au/aux/à Australia qui a été fabriqué par Rehab Assist Pty Ltd.

Qu'est-ce que c'est?

Une correction ou un retrait opéré par un fabricant afin de répondre à un problème causé par un dispositif médical. Les rappels surviennent lorsqu'un dispositif médical est défectueux, lorsqu'il pourrait poser un risque pour la santé, ou les deux à la fois.

En savoir plus sur les données ici
  • Type d'événement
    Recall
  • ID de l'événement
    RC-2016-RN-01200-1
  • Classe de risque de l'événement
    Class II
  • Date de mise en oeuvre de l'événement
    2016-09-14
  • Pays de l'événement
  • Source de l'événement
    DHTGA
  • URL de la source de l'événement
  • Notes / Alertes
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Notes supplémentaires dans les données
  • Cause
    Some c-series carry bars are attached to the track using a plastic swivel. if the plastic swivel in these bars becomes damaged there is the potential for it to separate from the carry bar, resulting in a patient fall. rehab assist will supply a free metal insert to be placed inside affected carry bars to prevent separation if the plastic swivel becomes damaged.There have been nine incidents worldwide in which the plastic swivel mechanism has been reported as damaged. no incidents have been reported in australia and there have been no injuries reported as a result of this issue.
  • Action
    Rehab Assist is advising distributors to contact all end users and inspect the swivels to determine if there are any damaged swivels in use. Customers are advised to inspect the swivel for damage or contact their distributor to arrange for a technician to do an inspection. If there is no damage the lift can continue to be used as normal. The metal insert should be installed as soon as practicable. If the swivel appears damaged, do not use the lift until the metal insert has been inserted. Wherever the plastic swivel is damaged, users are also advised to purchase and replace the carry bar as soon as possible. For more details, please see https://www.tga.gov.au/alert/rehab-assist-c-series-ceiling-hoists .

Device

  • Modèle / numéro de série
    C-series carry bars (used with C300, C450 Manual, C625 ceiling lifts/hoists)Models Affected: 360741, 360750, 360751, 360755, 360755 Rev B, 360756, 360757 (identifiable by a black plastic swivel/puck)ARTG Number: 169639