Rappel de Cios Alpha mobile C-arm x-ray system

Selon Department of Health, Therapeutic Goods Administration, ce/cet/cette rappel concerne un dispositif en/au/aux/à Australia qui a été fabriqué par Siemens Ltd.

Qu'est-ce que c'est?

Une correction ou un retrait opéré par un fabricant afin de répondre à un problème causé par un dispositif médical. Les rappels surviennent lorsqu'un dispositif médical est défectueux, lorsqu'il pourrait poser un risque pour la santé, ou les deux à la fois.

En savoir plus sur les données ici
  • Type d'événement
    Recall
  • ID de l'événement
    RC-2015-RN-00399-1
  • Classe de risque de l'événement
    Class II
  • Date de mise en oeuvre de l'événement
    2015-05-13
  • Pays de l'événement
  • Source de l'événement
    DHTGA
  • URL de la source de l'événement
  • Notes / Alertes
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Notes supplémentaires dans les données
  • Cause
    Under certain circumstances the following user actions may cause the system to freeze. ·when images were transferred from the left monitor to the right monitor (function a to b) during image acquisition (i.E. radiation on) or shortly before starting image acquisition·when acquisition of image was started while storing of scenes initiated by pressing the store button was still in progress·when scene replay was initiated either automatically or by pressing the “play” button (especially when pressing the “play” button more than once) while the storing of those scenes was still in progress·when image acquisition was initiated again before automatic replay of the last scene started·when finalized cd/dvds were used to store additional data·when radiation summary reports with more than 190 entries were being exported·when system restart was initiated when the error message “system blocked. turn system off/on” (error codes 215/80 and/or 208/80) was displayed.
  • Action
    Siemens has provided a workaround whilst a system modification is being developed which is expected for July 2015. This action has been closed-out on 11/08/2016.

Device

  • Modèle / numéro de série
    Cios Alpha mobile C-arm x-ray systemCatalogue number: 10308191ARTG number: 102180
  • Manufacturer

Manufacturer

  • Société-mère du fabricant (2017)
  • Source
    DHTGA