Rappel de Colleague Volumetric Infusion Pumps

Selon Department of Health, Therapeutic Goods Administration, ce/cet/cette rappel concerne un dispositif en/au/aux/à Australia qui a été fabriqué par Baxter Healthcare Pty Ltd.

Qu'est-ce que c'est?

Une correction ou un retrait opéré par un fabricant afin de répondre à un problème causé par un dispositif médical. Les rappels surviennent lorsqu'un dispositif médical est défectueux, lorsqu'il pourrait poser un risque pour la santé, ou les deux à la fois.

En savoir plus sur les données ici
  • Type d'événement
    Recall
  • ID de l'événement
    RC-2014-RN-00384-1
  • Classe de risque de l'événement
    Class II
  • Date de mise en oeuvre de l'événement
    2014-03-31
  • Pays de l'événement
  • Source de l'événement
    DHTGA
  • URL de la source de l'événement
  • Notes / Alertes
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Notes supplémentaires dans les données
  • Cause
    During 2011, it was identified that a quantity of pumps fitted with two ceramic 1.6a slow-blow fuses in the rear housing was failing prematurely. investigation led to a change of the fuse type fitted during future manufacture and service, from a ceramic fuse to a glass fuse, which has a different rupture characteristic. the failure rate of this reported issue has subsequently declined following the replacement with a glass fuse in newly manufactured pumps. if the fuses fail the device cannot charge the batteries even when connected to a.C. power. failure of the fuses while the pump is in operation would result in the pump switching from a.C. power to battery power and the a.C. icon would extinguish. the pump would continue to run on battery power with corresponding battery discharge alarms.
  • Action
    Customers are asked to examine the two fuses installed in the rear of the pump which can be accessed by slotted fuse covers to establish if the fuses are glass or ceramic type. Ceramic fuses should be replaced with the glass fuses which will be supplied by Baxter. For Baxter owned pumps placed with customers, Baxter will carry out this activity at the next service.

Device

  • Modèle / numéro de série
    Colleague Volumetric Infusion Pumps Product Codes: 2M91617 & 2M91637 ARTG Number: 177494