Rappel de Essenta DR Compact (General purpose X-ray system)

Selon Department of Health, Therapeutic Goods Administration, ce/cet/cette rappel concerne un dispositif en/au/aux/à Australia qui a été fabriqué par Philips Electronics Australia Ltd.

Qu'est-ce que c'est?

Une correction ou un retrait opéré par un fabricant afin de répondre à un problème causé par un dispositif médical. Les rappels surviennent lorsqu'un dispositif médical est défectueux, lorsqu'il pourrait poser un risque pour la santé, ou les deux à la fois.

En savoir plus sur les données ici
  • Type d'événement
    Recall
  • ID de l'événement
    RC-2012-RN-00787-3
  • Classe de risque de l'événement
    Class II
  • Date de mise en oeuvre de l'événement
    2012-08-03
  • Pays de l'événement
  • Source de l'événement
    DHTGA
  • URL de la source de l'événement
  • Notes / Alertes
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Notes supplémentaires dans les données
  • Cause
    Three issues requiring correction have been identified by the manufacturer:1. during reliability tests on the steel rope, one of the springs broke and the safety catch did not hold. the safety catch mechanism does not activate due to the counteracting force of the remaining spring. as a consequence the u-arm moves down uncontrolled. the manufacturer has identified that this can occur in the field earliest after 2.5 years of operation. these 2.5 years are calculated for a worst case usage of the system. 2. during installation all 3 screws for one sib chain wheel were found loose. if all 3 screws of one of these wheels loosen completely, the detector carriage and/or the tube carriage can move down uncontrolled until stopped by the corresponding end stop.3. during in-house testing it was detected that the life time of the wire rope is significantly lower than the specified 10 years. the replacement of the wire rope every 3 years is required.
  • Action
    Philips is urging customers to contact them if they have any concerns and is replacing the vertical carriage, wire rope, SID chain wheels and Instructions for Use.

Device

  • Modèle / numéro de série
    Essenta DR Compact (General purpose X-ray system)System Number: 9890-010-88201ARTG Number: 117662
  • Manufacturer

Manufacturer

  • Société-mère du fabricant (2017)
  • Source
    DHTGA