Rappel de Immulite 2000 / Immulite 2000 XPi Sex Hormone Binding Globulin (SHBG). An in vitro diagnostic medical device (IVD).

Selon Department of Health, Therapeutic Goods Administration, ce/cet/cette rappel concerne un dispositif en/au/aux/à Australia qui a été fabriqué par Siemens Healthcare Pty Ltd.

Qu'est-ce que c'est?

Une correction ou un retrait opéré par un fabricant afin de répondre à un problème causé par un dispositif médical. Les rappels surviennent lorsqu'un dispositif médical est défectueux, lorsqu'il pourrait poser un risque pour la santé, ou les deux à la fois.

En savoir plus sur les données ici
  • Type d'événement
    Recall
  • ID de l'événement
    RC-2017-RN-00196-1
  • Classe de risque de l'événement
    Class II
  • Date de mise en oeuvre de l'événement
    2017-02-08
  • Pays de l'événement
  • Source de l'événement
    DHTGA
  • URL de la source de l'événement
  • Notes / Alertes
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Notes supplémentaires dans les données
  • Cause
    Siemens healthcare diagnostics has confirmed that the immulite 2000/immulite 2000 xpi sex hormone binding globulin (shbg) adjustor lots 124 and 125 packed with specific reagent lots do not meet the thirty (30) day open vial adjustor stability claim when reconstituted and stored at 2-8°c, as published in the instructions for use. reconstituted adjustors stored at 2-8°c after a two week (14 days) period may show a bias in patient samples and may result in quality control falling outside of established ranges. the reagent lots affected can exhibit an average bias of 14.5% at day 21 to 16.5% at day 28 (for shbg concentrations of 5 nmol/l to 170 nmol/l) during the open vial adjustor stability period of 30 days. the bias observed was not concentration dependent. this issue does not impact the immulite/immulite1000 platforms.
  • Action
    Siemens is advising that reconstituted adjustor lots 124 and 125 are to be stored at 2-8°C or -20°C for up to 14 days, after which they must be discarded. Based on the recommended calibration interval, this allows two calibration events at day 1 and 14 after opening the SHBG reagent kit. Replacement product can be requested from Siemens if adjustors available in the laboratory are not sufficient to utilise the full reagent kit. The customer letter should be reviewed by the laboratory Medical Director and review of previously generated results is at the discretion of the laboratory.

Device

  • Modèle / numéro de série
    Immulite 2000 / Immulite 2000 XPi Sex Hormone Binding Globulin (SHBG). An in vitro diagnostic medical device (IVD).Catalogue Number: L2KSH2Siemens Material Number: 10381198Lot Numbers: 357 (Exp. 30 Jan 2017), 358 (Exp. 31 Mar 2017), 360 (Exp. 30 Apr 2017), 361 (Exp. 31 May 2017), 363 (Exp. 30 Jun 2017)ARTG number: 179720
  • Manufacturer

Manufacturer

  • Société-mère du fabricant (2017)
  • Source
    DHTGA