Rappel de Ingenuity TF PET/CT System with Addonics Mass Storage Device (MSD)

Selon Department of Health, Therapeutic Goods Administration, ce/cet/cette rappel concerne un dispositif en/au/aux/à Australia qui a été fabriqué par Philips Electronics Australia Ltd.

Qu'est-ce que c'est?

Une correction ou un retrait opéré par un fabricant afin de répondre à un problème causé par un dispositif médical. Les rappels surviennent lorsqu'un dispositif médical est défectueux, lorsqu'il pourrait poser un risque pour la santé, ou les deux à la fois.

En savoir plus sur les données ici
  • Type d'événement
    Recall
  • ID de l'événement
    RC-2018-RN-00233-1
  • Classe de risque de l'événement
    Class II
  • Date de mise en oeuvre de l'événement
    2018-03-15
  • Pays de l'événement
  • Source de l'événement
    DHTGA
  • URL de la source de l'événement
  • Notes / Alertes
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Notes supplémentaires dans les données
  • Cause
    Philips has identified that there may be an issue with acquisition if the addonics mass storage device (msd) provided with the system is connected to the console and powered on during an acquisition.Following a period of inactivity, the addonics msd may cause the acquisition console to become unresponsive. this may prevent an acquisition from proceeding resulting in a potential re-scan being required.
  • Action
    Philips is advising users to not view studies or access data on the Addonics MSD while a patient is being scanned. If a user is attempting to view information stored on the Addonics MSD and the console appears to be unresponsive, reset power on the Addonics MSD or pull out and reconnect its USB cable. It is recommended that the Addonics USB Mass Storage Device be turned off when not being accessed, and powered on only when data transfer is needed. Following data transfer, the Addonics USB Mass Storage Device should be turned off. Philips will be providing an updated Mass Storage Device to all affected customer to correct the issue.

Device

  • Modèle / numéro de série
    Ingenuity TF PET/CT System with Addonics Mass Storage Device (MSD)PET/CT System Model Number: 882442MSD Serial Number: 2090ARTG Number: 292543(Philips Electronics Australia - Combination diagnostic imaging system, PET/CT)
  • Manufacturer

Manufacturer

  • Société-mère du fabricant (2017)
  • Source
    DHTGA