Rappel de IntelliSpace PACS 4.4 (Picture archiving and communication system)

Selon Department of Health, Therapeutic Goods Administration, ce/cet/cette rappel concerne un dispositif en/au/aux/à Australia qui a été fabriqué par Philips Electronics Australia Ltd.

Qu'est-ce que c'est?

Une correction ou un retrait opéré par un fabricant afin de répondre à un problème causé par un dispositif médical. Les rappels surviennent lorsqu'un dispositif médical est défectueux, lorsqu'il pourrait poser un risque pour la santé, ou les deux à la fois.

En savoir plus sur les données ici
  • Type d'événement
    Recall
  • ID de l'événement
    RC-2013-RN-01072-1
  • Classe de risque de l'événement
    Class I
  • Date de mise en oeuvre de l'événement
    2013-10-17
  • Pays de l'événement
  • Source de l'événement
    DHTGA
  • URL de la source de l'événement
  • Notes / Alertes
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Notes supplémentaires dans les données
  • Cause
    Philips healthcare has become aware of a potential problem with the philips intellispace pacs system. there is a software issue which in certain circumstances could result in the unintended deletion of images sent to the intellispace pacs 4.4 before they are available for access by system users. timing of images being received by the pacs varies based on study size or modality configuration. if additional images are received into the study input folder as the input folder is being deleted, those images are not saved and can result in missing images.
  • Action
    Philips is providing temporary workaround instructions to mitigate the failure mode occurring. Philips is developing a software update that corrects this issue. It is anticipated this update will be available before the end of October 2013.