Rappel de iPlan RT, iPlan RT Dose and BrainSCANAll Software versions

Selon Department of Health, Therapeutic Goods Administration, ce/cet/cette rappel concerne un dispositif en/au/aux/à Australia qui a été fabriqué par Brainlab Australia Pty Ltd.

Qu'est-ce que c'est?

Une correction ou un retrait opéré par un fabricant afin de répondre à un problème causé par un dispositif médical. Les rappels surviennent lorsqu'un dispositif médical est défectueux, lorsqu'il pourrait poser un risque pour la santé, ou les deux à la fois.

En savoir plus sur les données ici
  • Type d'événement
    Recall
  • ID de l'événement
    RC-2016-RN-01641-1
  • Classe de risque de l'événement
    Class I
  • Date de mise en oeuvre de l'événement
    2016-12-23
  • Pays de l'événement
  • Source de l'événement
    DHTGA
  • URL de la source de l'événement
  • Notes / Alertes
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Notes supplémentaires dans les données
  • Cause
    Brainlab has identified an issue where there is the possibility of a potential inaccurate dose calculation for small mlc-shaped fields when using the pencil beam algorithm. the pencil beam algorithm refers to beam data which is acquired by performing dose measurements for mlc fields at predefined field sizes and which are stored using the iplan rt physics administration / brainscan beam profile editor. due to an inaccurate implementation in the brainlab radiation treatment planning software, the interpolation between two measured values is less accurate a) the smaller the field and;b) the further away the equivalent field size is from the measured value. for certain treatment setups, this leads to the effect that an insufficient amount of monitor units (mu) is calculated and that consequently the delivered dose is lower than the planned dose. for clarity, the delivered dose does not exceed the planned dose due to the described issue.
  • Action
    As a workaround, to reduce the described dose deviation to less than 2% using the Pencil Beam algorithm in iPlan RT 4.5.x Brainlab recommends to modify the machine profile following the instructions outlined in the Appendix provided. For all Brainlab treatment planning systems, as recommended by Brainlab, perform proper pre-treatment quality assurance for every treatment plan. Independent calculation of the dose using 3rd party QA Software or dose measurements inside a phantom is suitable to judge the accuracy of the dose calculation. Refer to the Brainlab Physics Technical Reference Guide for details.

Device

  • Modèle / numéro de série
    iPlan RT, iPlan RT Dose and BrainSCANAll Software versionsARTG Number: 121114
  • Manufacturer

Manufacturer