Rappel de Maquet Operating Table with Fixture 1002.65AO Manufactured in the period between 2009 and 2013

Selon Department of Health, Therapeutic Goods Administration, ce/cet/cette rappel concerne un dispositif en/au/aux/à Australia qui a été fabriqué par Maquet Australia Pty Ltd.

Qu'est-ce que c'est?

Une correction ou un retrait opéré par un fabricant afin de répondre à un problème causé par un dispositif médical. Les rappels surviennent lorsqu'un dispositif médical est défectueux, lorsqu'il pourrait poser un risque pour la santé, ou les deux à la fois.

En savoir plus sur les données ici
  • Type d'événement
    Recall
  • ID de l'événement
    RC-2016-RN-00984-1
  • Classe de risque de l'événement
    Class I
  • Date de mise en oeuvre de l'événement
    2016-07-22
  • Pays de l'événement
  • Source de l'événement
    DHTGA
  • URL de la source de l'événement
  • Notes / Alertes
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Notes supplémentaires dans les données
  • Cause
    The fixture (1002.65ao) is designed as a connecting element for the support and positioning of the patient’s head on a maquet operating table immediately before, during and after surgical interventions and as for examination and treatment. the fixture is used to adapt a range of different head rest and horseshoe head rests with a central square mount. post market surveillance reported that the welded joint between the square mount and the first fixable swivel joint had broken. as a cantilever design, this joint must bear the highest load of all welded joints. it has been determined that the breakage had been caused by ineffective welding in addition to rough handling and multiple overloading. serious injury to the patient's cervical spine may occur if the weld seam breaks during a procedure. there have been no reports of injuries as a result of this issue.
  • Action
    A Maquet service technician will be contacting customers to arrange an appointment to carry out replacement of all associated components. Until the replacement has been performed, users are advised that the device should be inspected for damage before every use. In this context, particular attention should be paid to the potential for cracks at the weld in the area illustrated in the customer letter. In case of doubt, customers are advised not to use the fixture. This action has been closed-out on 26/05/2017.