Rappel de Merge iConnect Enterprise Archive (iCEA) with software versions: 5.30.0 and greater up to and including 9.4.3(a picture archiving and communications system)

Selon Department of Health, Therapeutic Goods Administration, ce/cet/cette rappel concerne un dispositif en/au/aux/à Australia qui a été fabriqué par Emergo Asia Pacific Pty Ltd T/a Emergo Australia.

Qu'est-ce que c'est?

Une correction ou un retrait opéré par un fabricant afin de répondre à un problème causé par un dispositif médical. Les rappels surviennent lorsqu'un dispositif médical est défectueux, lorsqu'il pourrait poser un risque pour la santé, ou les deux à la fois.

En savoir plus sur les données ici
  • Type d'événement
    Recall
  • ID de l'événement
    RC-2016-RN-00873-1
  • Classe de risque de l'événement
    Class II
  • Date de mise en oeuvre de l'événement
    2016-06-30
  • Pays de l'événement
  • Source de l'événement
    DHTGA
  • URL de la source de l'événement
  • Notes / Alertes
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Notes supplémentaires dans les données
  • Cause
    This recall has been initiated because the impacted versions of icea allow images to be stored to the icea without pixel data. - image was retrieved by radsuite from icea, a qc action was performed, & sent back to icea. - image was sent to icea with pixel data but the pixel data was not stored with the image to icea. the investigation found that there is a networking scenario where the network dropped the connection after sending the header information but before the pixel data arrived. icea stored the image without the pixel data.- image was sent to icea without pixel data & the image was stored to icea. the pixel data may have been lost by the device that sent the data to icea.A side effect of images without pixel data is performance degradation caused by the system continuously attempting to route the images to the peer system that fail to store. there could be delay in treatment in determining the progression of disease.
  • Action
    Emergo is advising users that there is no work around for the issue, however the error is obvious to users using PACS viewer, as it is indicated that there is a missing image (slice) from the study. Users not using PACS viewer may not be notified of the error. Emergo will be implementing a software upgrade as a permanent correction. This action has been closed-out on 22/02/2017.

Device

  • Modèle / numéro de série
    Merge iConnect Enterprise Archive (iCEA) with software versions: 5.30.0 and greater up to and including 9.4.3(a picture archiving and communications system)ARTG Number: 206683