Rappel de MOSAIQ Medical Oncology version 2.50.05 and higher (oncology information system used to manage workflows for treatment planning and delivery)

Selon Department of Health, Therapeutic Goods Administration, ce/cet/cette rappel concerne un dispositif en/au/aux/à Australia qui a été fabriqué par Elekta Pty Ltd.

Qu'est-ce que c'est?

Une correction ou un retrait opéré par un fabricant afin de répondre à un problème causé par un dispositif médical. Les rappels surviennent lorsqu'un dispositif médical est défectueux, lorsqu'il pourrait poser un risque pour la santé, ou les deux à la fois.

En savoir plus sur les données ici
  • Type d'événement
    Recall
  • ID de l'événement
    RC-2016-RN-00954-1
  • Classe de risque de l'événement
    Class II
  • Date de mise en oeuvre de l'événement
    2016-07-20
  • Pays de l'événement
  • Source de l'événement
    DHTGA
  • URL de la source de l'événement
  • Notes / Alertes
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Notes supplémentaires dans les données
  • Cause
    It is possible that a change to an order set will not be saved in the current, open care plan when the following workflow is utilised: 1. open either the care plan edit form or the care plan instance form and leave the form open. 2. select an order set that is part of the same care plan and make a change to the order set. this can be a change to the order set name or label or a change to the drug, dose, administration instructions, etc. 3. the update made to the order set will not be saved in the current, open care plan.Although drug and/or dose information could be edited, but if substantial changes were made to drug information a new order set would be generated rather than editing an existing order set. it is more likely that the edits made would result in non-serious injury (such as administration details, order set name, etc.) there have been no injuries reported as a result of this issue.
  • Action
    Elekta is advising users to ensure all Care Plan windows are closed prior to making and updates to Order Sets. A software correction will be implemented as a permanent correction. This action has been closed-out on 24/05/2017.

Device

  • Modèle / numéro de série
    MOSAIQ Medical Oncology version 2.50.05 and higher (oncology information system used to manage workflows for treatment planning and delivery)ARTG Number: 225164