Rappel de Philips Model EPIQ 7 Ultrasound System with Pediatric Cardiology option(Intended for diagnostic ultrasound imaging and fluid flow analysis)

Selon Department of Health, Therapeutic Goods Administration, ce/cet/cette rappel concerne un dispositif en/au/aux/à Australia qui a été fabriqué par Philips Electronics Australia Ltd.

Qu'est-ce que c'est?

Une correction ou un retrait opéré par un fabricant afin de répondre à un problème causé par un dispositif médical. Les rappels surviennent lorsqu'un dispositif médical est défectueux, lorsqu'il pourrait poser un risque pour la santé, ou les deux à la fois.

En savoir plus sur les données ici
  • Type d'événement
    Recall
  • ID de l'événement
    RC-2015-RN-00407-1
  • Classe de risque de l'événement
    Class II
  • Date de mise en oeuvre de l'événement
    2015-05-15
  • Pays de l'événement
  • Source de l'événement
    DHTGA
  • URL de la source de l'événement
  • Notes / Alertes
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Notes supplémentaires dans les données
  • Cause
    When epiq 7 ultrasound system is set up to ‘metric’ and weight and/or height is entered, a unit conversion error may result in the incorrect calculation of the patient’s body surface area (bsa). using patient’s weight only to calculate the bsa, which is the usual practice for patients under the age of 2 years, would result in a value that is too high by a factor of about 1.7. a cardiac index calculated using that value would be too low by a factor of 1.7. an error of this magnitude may not be readily detected by the user.Alternatively, when using weight only the bsa may be out by a factor of about 1.7 which is much harder to detect as a cardiac index calculated using that value may not be as obvious to the user. however it is highly unlikely that this type of calculation would be used for patients older than 2 years of age. therefore, the risk of this error occurring is greatest when the bsa is calculated using weight only in patients under 2 years of age.
  • Action
    Philips has corrected this issue in software version 1.3.3. which will be provided to users free of charge. This action has been closed-out on 15/08/2016.

Device

  • Modèle / numéro de série
    Philips Model EPIQ 7 Ultrasound System with Pediatric Cardiology option(Intended for diagnostic ultrasound imaging and fluid flow analysis)Software versions 1.3.2 or lower ARTG Number: 93851