Rappel de QuikFlap Neuro Implants

Selon Department of Health, Therapeutic Goods Administration, ce/cet/cette rappel concerne un dispositif en/au/aux/à Australia qui a été fabriqué par Stryker Australia Pty Ltd.

Qu'est-ce que c'est?

Une correction ou un retrait opéré par un fabricant afin de répondre à un problème causé par un dispositif médical. Les rappels surviennent lorsqu'un dispositif médical est défectueux, lorsqu'il pourrait poser un risque pour la santé, ou les deux à la fois.

En savoir plus sur les données ici
  • Type d'événement
    Recall
  • ID de l'événement
    RC-2018-RN-00682-1
  • Classe de risque de l'événement
    Class II
  • Date de mise en oeuvre de l'événement
    2018-06-18
  • Pays de l'événement
  • Source de l'événement
    DHTGA
  • URL de la source de l'événement
  • Notes / Alertes
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Notes supplémentaires dans les données
  • Cause
    The current packaging and ifu indicate a five-year shelf life. during long-term product testing, it was discovered that after a three-year term, there is a potential for the external packaging (sterile barrier) to become compromised. such compromised packaging results in the loss of the vacuum seal allowing for contamination of the inside peel pouch. a breach of the outer peel pouch has only been observed with packages at shelf life end and after excessive transportation testing.To date, medtronic have not received any complaints as a result of this issue.
  • Action
    Stryker is advising customers to quarantine affected stock to prevent further use. Stryker representatives will contact customers to help coordinate the return of any affected product. In the event this device has been recently implanted, surgeons are advised to monitor for signs of infection as per their normal processes.

Device

  • Modèle / numéro de série
    QuikFlap Neuro ImplantsMultiple Item NumbersMultiple Lot Numbers ARTG Number: 295882(Stryker Australia - Cranofacial fixation plate, non-biodegradable)
  • Manufacturer

Manufacturer

  • Société-mère du fabricant (2017)
  • Source
    DHTGA