Rappel de Ready-Prep %CDT by HPLC using the CDM software version MVC133091. An in vitro diagnostic medical device (IVD)

Selon Department of Health, Therapeutic Goods Administration, ce/cet/cette rappel concerne un dispositif en/au/aux/à Australia qui a été fabriqué par Bio-Rad Laboratories Pty Ltd.

Qu'est-ce que c'est?

Une correction ou un retrait opéré par un fabricant afin de répondre à un problème causé par un dispositif médical. Les rappels surviennent lorsqu'un dispositif médical est défectueux, lorsqu'il pourrait poser un risque pour la santé, ou les deux à la fois.

En savoir plus sur les données ici
  • Type d'événement
    Recall
  • ID de l'événement
    RC-2017-RN-01532-1
  • Classe de risque de l'événement
    Class III
  • Date de mise en oeuvre de l'événement
    2017-12-20
  • Pays de l'événement
  • Source de l'événement
    DHTGA
  • URL de la source de l'événement
  • Notes / Alertes
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Notes supplémentaires dans les données
  • Cause
    Bio-rad have observed that the lot numbers for mobile phase 2 [mp][2] and mobile phase 3 [mp][3] on the cd-rom cover print and within the method for the cdm software are not correct for the affetced cd-rom, due to a defect in the manufacturing process of the cd-rom.Thus, the printed numbers on the cd-rom do not correspond to the correct lot numbers of mobile phase 2 and mobile phase 3 which were delivered with the associated kits.The wrong lot numbers within the cdm method will affect only customers using the cdm software: the wrong lot numbers within the method of the cdm software will result in transfer of the wrong lot numbers into the users cdm software, and false lot numbers printed in the summary reports of the analytical runs. this may result in difficulties of the traceability of used mobile phases. due to the fact that the wrong lot numbers are not used for any other reagents or components of any %cdt by hplc test there is no risk of false traceability.
  • Action
    Bio-Rad is advising users to immediately check the documentation obtained by using the CD-ROM of the affected lots. If users have used the CD-ROM to read in the lot numbers into the CDM software, correct the MP2 and 3 lot numbers in the CDM software manually (as described in the CDM manual Version 4.0, chapter 5.1.4). Alternatively, discontinue use of the CD-ROM and Bio-Rad will replace the CD-ROM when the quality issue is resolved.

Device

  • Modèle / numéro de série
    Ready-Prep %CDT by HPLC using the CDM software version MVC133091. An in vitro diagnostic medical device (IVD)Catalogue Number: 1956670, 1956671Serial numbers: 64121977, 64139856, 64139684, 64142119, 64121979Expiry Date: 15-03-2018ARTG Number: 194835