Rappel de Samarys Peek Anterior Cervical Cage

Selon Department of Health, Therapeutic Goods Administration, ce/cet/cette rappel concerne un dispositif en/au/aux/à Australia qui a été fabriqué par B Braun Australia Pty Ltd.

Qu'est-ce que c'est?

Une correction ou un retrait opéré par un fabricant afin de répondre à un problème causé par un dispositif médical. Les rappels surviennent lorsqu'un dispositif médical est défectueux, lorsqu'il pourrait poser un risque pour la santé, ou les deux à la fois.

En savoir plus sur les données ici
  • Type d'événement
    Recall
  • ID de l'événement
    RC-2015-RN-01246-1
  • Classe de risque de l'événement
    Class I
  • Date de mise en oeuvre de l'événement
    2015-12-21
  • Pays de l'événement
  • Source de l'événement
    DHTGA
  • URL de la source de l'événement
  • Notes / Alertes
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Notes supplémentaires dans les données
  • Cause
    This hazard alert is being issued because of a random non-conformity relevant to the gold wires used as a position marker for the cervical peek cage samarys. the observed facts were that some devices within 4 lots manufactured between april 2014 and november 2015 presented missing, improperly secured, and/or incorrectly positioned gold wire markers, which may protrude 1-2 mm inside or outside the cage.Gold wires protruding from cage may cause local inflammation and/or neural or vascular injuries at treated site or distant ones by migration (in complete protrusion), which potentially can have serious consequences. the probability of these potential adverse events is however very low. to date, no customer complaint has been recorded and no adverse event or incident has been reported.
  • Action
    The sponsor is advising the surgeons that in the case of improper fixation of the gold wire, migration of these small dimensioned gold wires in the surrounding tissues would be easily detected on the X-ray films that are routinely taken during patient follow-up. Additionally, as exposed subjects may be often asymptomatic for a long period, surgeons must check on the routine X-ray films that position markers did not migrate. They may also warn implanted patients to report any late dysphagia, sensation of foreign bodies into their throat or suspect cough or different concerning symptoms, as this might allow early detecting a potential complication and solving it. For further information, please see http://www.tga.gov.au/alert/samarys-polyethylethylketone-anterior-cervical-cage This action has been closed-out on 30/08/2016.

Device

  • Modèle / numéro de série
    Samarys Peek Anterior Cervical CageUnits from the following batches of Samarys Peek Anterior Cervical Cage are affected.Part No: 11CC2A15-5S Lot No:23LJ-19196Part No: 11CC2A17-6S Lot No: 23KY+19029Part No: 11CC2AC17-5S Lot No: 247P+19029Part No: 11CC2AC17-6S Lot No: 248K+19029ARTG Number: 101789