Rappel de syngo Workflow SLR with versions VA31, VA32, VB10, or VB20A (TBC)(Radiology information system)

Selon Department of Health, Therapeutic Goods Administration, ce/cet/cette rappel concerne un dispositif en/au/aux/à Australia qui a été fabriqué par Siemens Ltd.

Qu'est-ce que c'est?

Une correction ou un retrait opéré par un fabricant afin de répondre à un problème causé par un dispositif médical. Les rappels surviennent lorsqu'un dispositif médical est défectueux, lorsqu'il pourrait poser un risque pour la santé, ou les deux à la fois.

En savoir plus sur les données ici
  • Type d'événement
    Recall
  • ID de l'événement
    RC-2015-RN-00443-1
  • Classe de risque de l'événement
    Class II
  • Date de mise en oeuvre de l'événement
    2015-05-28
  • Pays de l'événement
  • Source de l'événement
    DHTGA
  • URL de la source de l'événement
  • Notes / Alertes
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Notes supplémentaires dans les données
  • Cause
    While configured for the batching of orders, when an order transaction is received from a his interface, it is temporarily read into the system while waiting for the next transaction to come across the interface. if the next transaction is another procedure for the same patient, it would be "batched" with the first order until a transaction that is not an order for the "batched" patient is received or the maximum wait time of 10 seconds is exceeded. at this point, all the order(s) would be processed.If the system receives a shutdown request to restart the interface process while an order is within this 10 second "batching" window, the order may be lost with no indication to the user. this condition can happen when a user manually attempts to restart the interface process or when all of the application processes are restarted during the day-end processing sequence.
  • Action
    The issue can be avoided by not using the Order Batching feature. If the Order Batching feature is used. Users are recommended to configure the day-end processing to take place at the time with the anticipated lowest number of order transactions until the issue is solved in the software.For new upgrades or new installations, the issue will be solved in the syngo Workflow SLR software version VB20A_HF01 which is planned to be released in 02/2015. For existing VA31, VA32, VB10, and VB20A sites, the issue will be resolved following Safety Update Instruction SY038/15/S which is planned to be released in 02/2015. This action has been closed-out on 31/08/2016.

Device

  • Modèle / numéro de série
    syngo Workflow SLR with versions VA31, VA32, VB10, or VB20A (TBC)(Radiology information system) Catalogue number: 10558586ARTG number: 142663
  • Manufacturer

Manufacturer

  • Société-mère du fabricant (2017)
  • Source
    DHTGA