Rappel de Tandem Bipolar Hip System (hemiarthroplasty hip prosthesis)

Selon Department of Health, Therapeutic Goods Administration, ce/cet/cette rappel concerne un dispositif en/au/aux/à Australia qui a été fabriqué par Smith & Nephew Pty Ltd.

Qu'est-ce que c'est?

Une correction ou un retrait opéré par un fabricant afin de répondre à un problème causé par un dispositif médical. Les rappels surviennent lorsqu'un dispositif médical est défectueux, lorsqu'il pourrait poser un risque pour la santé, ou les deux à la fois.

En savoir plus sur les données ici
  • Type d'événement
    Recall
  • ID de l'événement
    RC-2016-RN-01167-1
  • Classe de risque de l'événement
    Class II
  • Date de mise en oeuvre de l'événement
    2016-09-06
  • Pays de l'événement
  • Source de l'événement
    DHTGA
  • URL de la source de l'événement
  • Notes / Alertes
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Notes supplémentaires dans les données
  • Cause
    Smith & nephew have identified a manufacturing error which affects certain batches. the affected devices were manufactured with a retainer groove that is out-of-specification. the out of specification condition could potentially result in one of three functional failures: the device will not assemble, the head is too tight or in rare occasions, the device may appear to be assembled properly during surgery but later disassociate post-operatively. the likelihood of implant detachment after surgery is very low. however, if it did occur, it would happen within the first few weeks after the operation and be clearly recognised by the patient (and confirmed by x-ray).
  • Action
    Smith & Nephew is advising users that there is no recommendation for proactive revisions. Failure of the device will result in the patient experiencing pain, limited mobility or loss of range-of-motion. Surgeons should maintain their routine follow-up protocol for patients who have been implanted with a potentially affected device. Smith & Nephew is advising users to inspect stock and quarantine the affected devices for return. Affected devices can be returned for credit or replaced with unaffected product. This action has been closed-out on 24/05/2017.

Device

  • Modèle / numéro de série
    Tandem Bipolar Hip System (hemiarthroplasty hip prosthesis)Material Numbers: 71324041, 71324044, 71324045, 71324048, 71324049, 71324050, 71324051, 71324052, 71324053, 71324054, 71324055, Multiple Batch Numbers affectedARTG Number: 242993
  • Manufacturer

Manufacturer