Rappel de V5Ms transesophageal transducer (used for ultrasound imaging)

Selon Department of Health, Therapeutic Goods Administration, ce/cet/cette rappel concerne un dispositif en/au/aux/à Australia qui a été fabriqué par Siemens Ltd.

Qu'est-ce que c'est?

Une correction ou un retrait opéré par un fabricant afin de répondre à un problème causé par un dispositif médical. Les rappels surviennent lorsqu'un dispositif médical est défectueux, lorsqu'il pourrait poser un risque pour la santé, ou les deux à la fois.

En savoir plus sur les données ici
  • Type d'événement
    Recall
  • ID de l'événement
    RC-2014-RN-00943-1
  • Classe de risque de l'événement
    Class I
  • Date de mise en oeuvre de l'événement
    2014-08-29
  • Pays de l'événement
  • Source de l'événement
    DHTGA
  • URL de la source de l'événement
  • Notes / Alertes
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Notes supplémentaires dans les données
  • Cause
    Seiemens has received recent overseas reports about the deterioration of material covering the articulating section of the v5ms transesophogeal transducer. although to date, no patient injury has been reported, a compromise or fault in this material combined with a failure to use a probe cover as directed in the user manual may cause oesophageal cuts, bleeding, and perforation to the patient, and severe trauma, electrical burns, and serious electrical hazards to the patient and the user. however, a leakage current test can indicate cracks, cuts, tears, perforations, and protrusions in the articulation material.
  • Action
    To avoid the potential risk associated with this issue, customers are asked to undertake the following steps: - Perform the leakage current test for a transesophageal transducer prior to each use. - Use only Siemens approved cleaning and disinfecting solutions. - Use market-cleared transducer sheaths specifically designed for TEE applications. - When users store the transesophageal transducer, take extreme care that the distal tip is straight. Never move the distal tip by hand. Always use the flex controls. Customers are advised that if they notice any damage or wear on the transducer, or if the transducer fails to pass the leakage current test, immediately discontinue use of the transducer and contact their local Customer Service Engineer. This action has been closed-out on 06/06/2016.