Alerte De Sécurité sur Commercial Names: CT Scanning System / Ingenuity CT; ANVISA registration numbers: 10216710191/10216710209; Risk class: III; Affected models: Brilliance 64, Ingenuity Core, Ingenuity Core128; Affected serial numbers: 10782, 10768, 10777, 10770, 10769, 95839, 95841, 95840, 95866, 95847, 95863, 95846, 95846, 320113.

Selon Agência Nacional de Vigilância Sanitária (ANVISA), ce/cet/cette alerte de sécurité concerne un dispositif en/au/aux/à Brazil qui a été fabriqué par Philips Medical Systems Ltda.; Philips Medical Systems (Cleveland), Inc. / Philips Medical Systems Ltda.

Qu'est-ce que c'est?

Les alertes fournissent des informations importantes et des recommandations concernant les dispositifs médicaux. Le fait qu'une alerte soit émise ne signifie pas nécessairement qu'un dispositif soit dangereux. Les alertes de sécurité, qui sont envoyées tant aux travailleurs du secteur médical qu'aux utilisateurs de ces dispositifs, peuvent inclure des rappels. Elles peuvent être rédigées par des fabricants mais aussi par des autorités en charge de la santé.

En savoir plus sur les données ici
  • Type d'événement
    Safety alert
  • ID de l'événement
    1892
  • Date
    2016-05-06
  • Pays de l'événement
  • Source de l'événement
    ANVISA
  • URL de la source de l'événement
  • Notes / Alertes
    Brazilian data is current through June 2018. All of the data comes from Anvisa, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Brazil.
  • Notes supplémentaires dans les données
    Philips Healthcare is implementing the software version 3.5.5 upgrade to correct the problem described above. Solution Center will contact you to schedule the software update installation on your site. If you need more information or support for this issue, please contact our Solution Center at 0800-701-7789. Service from Monday to Saturday from 07 to 19h.
  • Cause
    Philips healthcare received field reports stating that certain brilliance 64, ingenuity core and ingenuity core128 systems running software version 3.5.4 exhibited intermittent swirl-like circular artifacts that may appear on rebuilt images and may diagnosis of affected images. the clinician should apply clinical judgment to determine if the images contained in the data set without artifacts contain enough information to make a diagnosis. if these images are sufficient, no action is required. if the images are not sufficient, a new patient scan may be required.
  • Action
    Action code FCO72800597. Software update