Alerte De Sécurité sur DHS / DCS MEDIUM INSTRUMENT Technical Name: SURGICAL INSTRUMENTS ANVISA Registration Number: 80145901733 Hazard Class: I Affected Model: Instrumental Non-Articulated No Shearing Affected Series Numbers: SKU: 338.260 || Lot: All lot numbers <8016184 SKU: 338.280 || Lot: All lot numbers <7985313

Selon Agência Nacional de Vigilância Sanitária (ANVISA), ce/cet/cette alerte de sécurité concerne un dispositif en/au/aux/à Brazil qui a été fabriqué par Johnson & Johnson do Brasil Ind. e Com. de Prod. para Saúde Ltda; Synthes GmbH.

Qu'est-ce que c'est?

Les alertes fournissent des informations importantes et des recommandations concernant les dispositifs médicaux. Le fait qu'une alerte soit émise ne signifie pas nécessairement qu'un dispositif soit dangereux. Les alertes de sécurité, qui sont envoyées tant aux travailleurs du secteur médical qu'aux utilisateurs de ces dispositifs, peuvent inclure des rappels. Elles peuvent être rédigées par des fabricants mais aussi par des autorités en charge de la santé.

En savoir plus sur les données ici
  • Type d'événement
    Safety alert
  • ID de l'événement
    2108
  • Date
    2016-10-27
  • Pays de l'événement
  • Source de l'événement
    ANVISA
  • URL de la source de l'événement
  • Notes / Alertes
    Brazilian data is current through June 2018. All of the data comes from Anvisa, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Brazil.
  • Notes supplémentaires dans les données
    We have recorded that your facility has received the product (s) subject to this recall. It is important to note that withdrawal of the product would prevent emergency surgery, so Synthes GmbH is not requiring immediate removal of the affected product (s). Synthes GmbH is in the process of developing a plan for the recovery and replacement of the indicated devices and will contact you as soon as the replacement part becomes available.
  • Cause
    There is no evidence of the biocompatibility of the material of the dhs / dcs no. 338.280 impact insert, single (part number 338.260), used in the existing product (specified lots) (polyamide 6.6 with 20% reinforcement in carbon fiber, tecamid 66 cf 20 ). synthes has not performed biocompatibility tests for this material, and there is no evidence that it has been performed by other entities.
  • Action
    Field Action Code R2014204R triggered under the responsibility of Johnson & Johnson do Brasil Ind. And Com. De Prod. for Health Ltda. Company will make collection for later return to the manufacturer.

Manufacturer