Alerte De Sécurité sur Equipment for angiography ALLURA XPER, brand PHILIPS, Risk class: III, model: ALLURA XPER FD10 (floor), ALLURA XPER FD10 (ceiling), ALLURA XPER FD20 (floor), ALLURA XPER FD20 (ceiling), Registry: 10216710153; ///// ALLURA XPER ANGIOGRAPHY EQUIPMENT, model: ALLURA XPER FD10 / 10; ALLURA XPER FD20 / 10; ALLURA XPER FD20 / 20; ALLURA XPER FD 20/15; ALLURA XPER FD10 / 10 OR Table; ALLURA XPER FD20 / 10 OR Table; ALLURA XPER FD20 / 20 OR Table; ALLURA XPER FD 20/15 OR Table .. hazard class III, registration 10216710206. /// Serial numbers: 722006; 722124.

Selon Agência Nacional de Vigilância Sanitária (ANVISA), ce/cet/cette alerte de sécurité concerne un dispositif en/au/aux/à Brazil qui a été fabriqué par Philips Medical Systems Ltda.; Philips Medical Systems Nederland B.V..

Qu'est-ce que c'est?

Les alertes fournissent des informations importantes et des recommandations concernant les dispositifs médicaux. Le fait qu'une alerte soit émise ne signifie pas nécessairement qu'un dispositif soit dangereux. Les alertes de sécurité, qui sont envoyées tant aux travailleurs du secteur médical qu'aux utilisateurs de ces dispositifs, peuvent inclure des rappels. Elles peuvent être rédigées par des fabricants mais aussi par des autorités en charge de la santé.

En savoir plus sur les données ici
  • Type d'événement
    Safety alert
  • ID de l'événement
    1742
  • Date
    2015-11-09
  • Pays de l'événement
  • Source de l'événement
    ANVISA
  • URL de la source de l'événement
  • Notes / Alertes
    Brazilian data is current through June 2018. All of the data comes from Anvisa, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Brazil.
  • Notes supplémentaires dans les données
    According to the company, nonconformity does not directly cause a risky situation, however the beep is an available tool to help prevent unnecessary radiation on the patient. The field action application kits are still being prepared. #### UPDATED ON 07/18/2017, the company submitted a completion report of the field action proving the accomplishment of correction, as planned.
  • Cause
    Under certain circumstances, a software error may lead to a situation where the five-minute fluoroscopy beep is not emitted as required in 21cfr1020.32 (h) (2) (ii) and iec 60601-2-54 , clause 203.6.2.1.C. no injuries attributed to the problem were reported. it is an intermittent problem.
  • Action
    The field action addresses problem correction through software update. A change request is being sent in order to request installation of the R2.x version of the software that solves the problem in the alarm. According to manufacturer, the non-emission of the audible alarm occurs very intermittently. The user should always observe the real-time dose information and cumulative fluoroscopy time provided by the system. The fault condition is reset when the case of a new patient is started or when the system is reset. Action code: FCO 72200287