Alerte De Sécurité sur Name of the product: cobas® connection modules (CCM) - OUTPUT UNIT, accessory of the products under registration 10287411162 (cobas p512 and p612) Technical name: SAMPLE PROCESSOR ACCESSORY FOR HEALTH DIAGNOSIS ANVISA registration number: 10287411162 Hazard class: I Numbers Affected series: 20496176 204C5683 204C5684 204C5691 204C5692 204C5693 204C5694 204C5695 204C5700 (Roche stock) 204C5701 (Roche stock) 204C5702 (Roche stock)

Selon Agência Nacional de Vigilância Sanitária (ANVISA), ce/cet/cette alerte de sécurité concerne un dispositif en/au/aux/à Brazil qui a été fabriqué par Roche Diagnóstica Brasil Ltda; Roche Diagnostics GmBH.

Qu'est-ce que c'est?

Les alertes fournissent des informations importantes et des recommandations concernant les dispositifs médicaux. Le fait qu'une alerte soit émise ne signifie pas nécessairement qu'un dispositif soit dangereux. Les alertes de sécurité, qui sont envoyées tant aux travailleurs du secteur médical qu'aux utilisateurs de ces dispositifs, peuvent inclure des rappels. Elles peuvent être rédigées par des fabricants mais aussi par des autorités en charge de la santé.

En savoir plus sur les données ici
  • Type d'événement
    Safety alert
  • ID de l'événement
    2291
  • Date
    2017-05-17
  • Pays de l'événement
  • Source de l'événement
    ANVISA
  • URL de la source de l'événement
  • Notes / Alertes
    Brazilian data is current through June 2018. All of the data comes from Anvisa, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Brazil.
  • Notes supplémentaires dans les données
    If you identify that your unit is affected by rack vibration and samples are required for further processing, the unit should not be used until the Roche Diagnostica Brazil Field Service Representative performs the adjustment procedure. Potentially affected sample tubes should be disposed of in accordance with local regulations. An adjustment procedure that solves this problem is now available. All affected systems will be adjusted by a Field Service Representative from Roche Diagnostics Brazil.
  • Cause
    Roche diagnostics gmbh, a legal offshore manufacturer, has received a 5-position rack-related complaint on the output unit unit that was not being gently pushed into the tray and may cause rack vibration during sample transport, leading to possible spillage of the sample tube material.
  • Action
    Field Action Code SBN-CPS-2017-008 triggered under the responsibility of the company Roche Diagnóstica Brasil Ltda. Company will make correction in the field.