Alerte De Sécurité sur Non-rechargeable Lithium Batteries used with Lifepak 500 Automated External Defibrillators (Nonrechargeable Lithium Batteries used with Lifepak 500 Automated External Defibrillators). Identifier: Parts Nos .: (1) from 3005380-000 to 3005380-006, (2) 3005380-014; Date Codes: (1) from 9837 to 9920, (2) 9812, from 9901 to 9905 within Medtronic Physio-Control

Selon Agência Nacional de Vigilância Sanitária (ANVISA), ce/cet/cette alerte de sécurité concerne un dispositif en/au/aux/à Brazil qui a été fabriqué par Medtronic Physio-Control Corp.

Qu'est-ce que c'est?

Les alertes fournissent des informations importantes et des recommandations concernant les dispositifs médicaux. Le fait qu'une alerte soit émise ne signifie pas nécessairement qu'un dispositif soit dangereux. Les alertes de sécurité, qui sont envoyées tant aux travailleurs du secteur médical qu'aux utilisateurs de ces dispositifs, peuvent inclure des rappels. Elles peuvent être rédigées par des fabricants mais aussi par des autorités en charge de la santé.

En savoir plus sur les données ici
  • Type d'événement
    Safety alert
  • ID de l'événement
    283
  • Date
    2000-09-29
  • Pays de l'événement
  • Source de l'événement
    ANVISA
  • URL de la source de l'événement
  • Notes / Alertes
    Brazilian data is current through June 2018. All of the data comes from Anvisa, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Brazil.
  • Notes supplémentaires dans les données
    To date, the existence of 02 registrations of said product in Brazil on behalf of the Companies, UNITED MEDICAL LTDA and MEDTRONIC COMERCIAL LTDA, has been detected. BRASIL - 15.01.2002 - The Company UNITED MEDICAL LTDA, further sells said product, and is finalizing the process of requesting cancellation of registration with ANVISA. BRAZIL - 25.07.2002 - The company MEDTRONIC COMERCIAL LTDA, reports that 46 units were imported, and only 24 units belong to the lot / series that presented problems (30.01.2002). In addition, all replacements and upgrades have been successfully completed.
  • Cause
    Medtronic physio-control received 6 reports of battery failure caused by battery cell rupture. medtronic physio-control stated that each failure involved the rupture of only one battery which led to the opening / rupturing of the packaging box. two of the six reports said that this rupture caused minor injuries to the defibrillator operator, including respiratory irritations and first-degree burns. medtronic physio-control initiated a correction by correspondence dated july 10, 2000.
  • Action
    Make sure that you have received the July 10, 2000 correspondence and Medtronic Physio-Control replacement batteries. Identify the affected product by removing the battery from the Lifepak 500 automated external defibrillator (AED) to check the part number and date code. If the battery number does not match the numbers of the affected batteries, the battery can be reinstalled. If the battery number matches the affected battery numbers, remove it from use and install a new battery supplied by Medtronic Physio-Control. After installing the new battery, turn the AED on to advise the unit that a new battery has been installed. After 10 seconds, make sure the message "Connected Electrodes" appears. The unit is ready for use. The company estimates that the recall date for this recall is September 30, 2000. For more information, contact your Medtronic Physio-Control representative or technical support at (1-425) 867-4000 in the United States