Alerte De Sécurité sur Product Category: Angiography Equipment Allura Xper. Technical Name: Angiography Equipment. ANVISA registration number: 10216710153. Class of risk: III. Affected Model: Allura Xper FD10 Ceiling / Allura Xper FD10 Ceiling / Allura Xper FD10 Ceiling / Allura Xper FD10 Ceiling / Allura Xper FD10 Serial numbers affected: 947

Selon Agência Nacional de Vigilância Sanitária (ANVISA), ce/cet/cette alerte de sécurité concerne un dispositif en/au/aux/à Brazil qui a été fabriqué par Philips Medical Systems Ltda;: Philips Medical Systems Nederland B. V..

Qu'est-ce que c'est?

Les alertes fournissent des informations importantes et des recommandations concernant les dispositifs médicaux. Le fait qu'une alerte soit émise ne signifie pas nécessairement qu'un dispositif soit dangereux. Les alertes de sécurité, qui sont envoyées tant aux travailleurs du secteur médical qu'aux utilisateurs de ces dispositifs, peuvent inclure des rappels. Elles peuvent être rédigées par des fabricants mais aussi par des autorités en charge de la santé.

En savoir plus sur les données ici
  • Type d'événement
    Safety alert
  • ID de l'événement
    2547
  • Date
    2018-04-17
  • Pays de l'événement
  • Source de l'événement
    ANVISA
  • URL de la source de l'événement
  • Notes / Alertes
    Brazilian data is current through June 2018. All of the data comes from Anvisa, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Brazil.
  • Notes supplémentaires dans les données
    To prevent the problem from occurring, clients should reboot the system at least once a day. The customer shall ensure that all personnel accessing the affected systems are informed of the contents of the Security Notice. A copy of the Safety Notice must be placed with the system documentation until the system is corrected by Philips. If you wish to notify technical complaints and adverse events use the following channels: Notivisa: Notices of adverse events (AE) and technical complaints (QT) for products subject to Sanitary Surveillance should be made through the NOTIVISA System. To access the System, you must register and select the Health Professional option, if you are a liberal professional or the Institution / Entity option, if you are a professional of an institution / entity. Technovigilance System: Patient or citizen can notify through the Technovigilance System / SISTEC access through the link Additional information: - Date of identification of the problem by the company: 03/13/2018 - Date of notification notice to Anvisa: 03/21/2018 The company that owns the affected product is responsible for contacting its customers in a timely manner. to ensure the effectiveness of the Field Action in progress. It stands out the joint responsibility of the distribution chain and use of health products in maintaining its quality, safety and efficacy, as well as the effectiveness of the Field Action, expressed by RDC 23/2012: "(...) Art 2 ° A holder of registration of health product is the holder of the registration / registration of health product with Anvisa. Single paragraph. The registration holder, as well as other agents involved from the production to the use of the product, or discarding it when applicable, are jointly and severally responsible for maintaining the quality, safety and efficacy of the health products to the final consumer. Art. 12 Distributors of health products shall forward to the registration holder, in a timely manner, the distribution map and other information requested for the notification and execution of field actions. (...) "
  • Cause
    After continuous operation for a day and a half, the image on the flexvision large-screen monitor freezes for about 15 seconds. after this period, the system resumes and returns to normal operation. note: this problem applies only to allura xper and uniq systems with flexvision large-screen monitor. all other system monitors are not affected.
  • Action
    Field Action Code FCO72200422 triggered under the responsibility of the company Philips Medical Systems Ltda. Field correction.

Manufacturer

  • Société-mère du fabricant (2017)
  • Source
    ANVSANVISA