Alerte De Sécurité sur Product name: Acetabular capsule for hip arthroplasty Registration number ANVISA: 10417940030 Hazard class: III Model affected: Acceptable MD4 Affected batches: 02243Q; 12803P; 10652Q; 10654Q.

Selon Agência Nacional de Vigilância Sanitária (ANVISA), ce/cet/cette alerte de sécurité concerne un dispositif en/au/aux/à Brazil qui a été fabriqué par MDT Industria Comércio Importação e Exportação de Implantes S.A.; MDT Industria Comércio Importação e Exportação de Implantes S.A..

Qu'est-ce que c'est?

Les alertes fournissent des informations importantes et des recommandations concernant les dispositifs médicaux. Le fait qu'une alerte soit émise ne signifie pas nécessairement qu'un dispositif soit dangereux. Les alertes de sécurité, qui sont envoyées tant aux travailleurs du secteur médical qu'aux utilisateurs de ces dispositifs, peuvent inclure des rappels. Elles peuvent être rédigées par des fabricants mais aussi par des autorités en charge de la santé.

En savoir plus sur les données ici
  • Type d'événement
    Safety alert
  • ID de l'événement
    2219
  • Date
    2017-02-24
  • Pays de l'événement
  • Source de l'événement
    ANVISA
  • URL de la source de l'événement
  • Notes / Alertes
    Brazilian data is current through June 2018. All of the data comes from Anvisa, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Brazil.
  • Notes supplémentaires dans les données
    Non-deployed products should be segregated and returned to the MDT company. We recommend to users and patients that there are some technical complaints and adverse events involving the products mentioned in the Field Action (which can be verified on the Traceability Labels sent to each product), the Distributor Company and Faricante MDT and / or through the NOTIVISA System. To the users, it is confirmed that clinical follow-up of patients already implanted is not necessary, since possible failures are perceptible during the implantation procedure, making it possible to substitute the product and / or make the surgery unfeasible.
  • Cause
    Difficulty of coupling and / or unsuitable locking of the acetabular insert in the metallic shell.
  • Action
    Field Action Code AC 01.2017 triggered under the responsibility of the company MDT. Company will collect.