Alerte De Sécurité sur Trade Name: Articular Cannulas with MAQUET Reinforcement. Technical Name: Cannulae. ANVISA registration number: 10390690100. Risk Class: IV. Affected Model: Articular Cannulas with MAQUET Reinforcement. Serial numbers affected: A202107; A222107; A242107; ACR0890V; ACR1890V; AS18V; AS21V; AS24V

Selon Agência Nacional de Vigilância Sanitária (ANVISA), ce/cet/cette alerte de sécurité concerne un dispositif en/au/aux/à Brazil qui a été fabriqué par Maquet Cardiopulmonary do Brasil Industria e Comercio Ltda..

Qu'est-ce que c'est?

Les alertes fournissent des informations importantes et des recommandations concernant les dispositifs médicaux. Le fait qu'une alerte soit émise ne signifie pas nécessairement qu'un dispositif soit dangereux. Les alertes de sécurité, qui sont envoyées tant aux travailleurs du secteur médical qu'aux utilisateurs de ces dispositifs, peuvent inclure des rappels. Elles peuvent être rédigées par des fabricants mais aussi par des autorités en charge de la santé.

En savoir plus sur les données ici
  • Type d'événement
    Safety alert
  • ID de l'événement
    2585
  • Date
    2018-06-04
  • Pays de l'événement
  • Source de l'événement
    ANVISA
  • URL de la source de l'événement
  • Notes / Alertes
    Brazilian data is current through June 2018. All of the data comes from Anvisa, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Brazil.
  • Notes supplémentaires dans les données
    Recommendations (what to do, contacts with the manufacturer, how to notify Anvisa, etc.): Immediately return the affected products to the representative of the company holding the registration. If you wish to notify technical complaints and adverse events use the following channels: Notivisa: Notices of adverse events (EA) and technical complaints (QT) for products subject to Sanitary Surveillance should be made through the NOTIVISA System (http: //portal.anvisa .gov.br / notivisa). To access the System, you must register and select the Health Professional option, if you are a liberal professional or the Institution / Entity option, if you are a professional of an institution / entity. Technovigilance System: Patient or citizen can notify through the Technovigilance System / SISTEC access through the link Complementary Information: - Date of identification of the problem by the company: 05/02/2018 - Date of notification notice to Anvisa: 05/14/2018 The company that owns the affected product is responsible for contacting its customers in a timely manner. to ensure the effectiveness of the Field Action in progress. It stands out the joint responsibility of the distribution chain and use of health products in maintaining its quality, safety and efficacy, as well as the effectiveness of the Field Action, expressed by RDC 23/2012: "(...) Art 2 ° A holder of registration of health product is the holder of the registration / registration of health product with Anvisa. Single paragraph. The registration holder, as well as other agents involved from the production to the use of the product, or discarding it when applicable, are jointly and severally responsible for maintaining the quality, safety and efficacy of the health products to the final consumer. Art. 12 Distributors of health products shall forward to the registration holder, in a timely manner, the distribution map and other information requested for the notification and execution of field actions. (...) "
  • Cause
    An arterial cannula used for open heart surgery is usually equipped with a ventilation plug that facilitates cannula deaeration. the registration holder's airway cannula ventilation plug consists of sintered polyethylene, so the structure of the ventilation plug is porous and permeable by gas. after the deaeration process is completed, the client needs to separate the ventilation plug from the cannula to connect the extracorporeal circuit. the company states that in some cases, it was reported that the ventilation plug broke during separation of the arterial cannula (cases occurred outside brazil).
  • Action
    Field Action Code FSCA-2018-03-27 under the responsibility of Maquet Cardiopulmonary do Brasil Industria e Comercio Ltda. Gathering. Undoing.

Manufacturer

  • Société-mère du fabricant (2017)
  • Source
    ANVSANVISA