Alerte De Sécurité sur Trade name: DBS ELETRODE FOR DEEP BRAIN STIMULATION; DEEP BRAIN STIMULATION ELECTRODE; KIT CATETER DBS ELECTRODE FOR DEEP BRAIN STIMULATION Technical name: DBS ELETRODE FOR DEEP BRAIN STIMULATION; DEEP BRAIN STIMULATION ELECTRODE; KIT CATETER DBS ELECTRODE FOR DEEP BRAIN STIMULATION ANVISA registration number: 10339190484; 10339190549; 10339190339 Hazard Class: IV Affected Model: 3387S-40, 3389S-28, 3389S-40 and 3391S-40

Selon Agência Nacional de Vigilância Sanitária (ANVISA), ce/cet/cette alerte de sécurité concerne un dispositif en/au/aux/à Brazil qui a été fabriqué par MEDTRONIC COMERCIAL LTDA; Medtronic Inc..

Qu'est-ce que c'est?

Les alertes fournissent des informations importantes et des recommandations concernant les dispositifs médicaux. Le fait qu'une alerte soit émise ne signifie pas nécessairement qu'un dispositif soit dangereux. Les alertes de sécurité, qui sont envoyées tant aux travailleurs du secteur médical qu'aux utilisateurs de ces dispositifs, peuvent inclure des rappels. Elles peuvent être rédigées par des fabricants mais aussi par des autorités en charge de la santé.

En savoir plus sur les données ici
  • Type d'événement
    Safety alert
  • ID de l'événement
    2231
  • Date
    2017-03-17
  • Pays de l'événement
  • Source de l'événement
    ANVISA
  • URL de la source de l'événement
  • Notes / Alertes
    Brazilian data is current through June 2018. All of the data comes from Anvisa, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Brazil.
  • Notes supplémentaires dans les données
    If your surgical procedure involves the use of Medtronic's DBS depth limiter, Medtronic recommends the following: • For the implanted product: If the product has already been implanted, no action is necessary once the electrode placement at the target site was confirmed by intraoperative test stimulus, imaging and / or efficacy of the therapy. • For the product to be implanted: If it is suspected that the depth limiter has not been properly attached to the electrode, do not use and complete the procedure using a depth limiter from another Medtronic DBS electrode kit. As stated in the DBS electrode implant manual, Medtronic recommends checking the effect of the stimulus during the implant procedure and using imaging techniques to confirm electrode placement. #### UPDATED ON 7/24/2017, Medtronic Company submitted the field action completion report verifying the submission of the safety notice.
  • Cause
    The medtronic dbs depth limiter, also referred to as electrode holder, is used to establish the depth of the implant. the depth limiter interacts with parts of the stereotactic system used during surgery to control the placement of the tip of the dbs electrode at the target site. this letter provides information regarding a product performance problem that has been identified in this component, potential patient risks, and actions to be taken in connection with the use of medtronic's depth limiter.
  • Action
    Field Action Code FA761 triggered under the responsibility of the company MEDTRONIC COMERCIAL LTDA. Company will update, correct or supplement the instructions for use.

Manufacturer

  • Société-mère du fabricant (2017)
  • Commentaire du fabricant
    “If our surveillance systems identify a potential performance issue, our personnel promptly evaluate the problem, including, when appropriate, conducting root cause investigations and internal testing to assess whether the product continues to meet specifications and defined performance criteria,” Medtronic told ICIJ in a statement. “In some cases, based on this evaluation, Medtronic may determine that a recall is necessary.” The company said that it communicates with healthcare providers and/or patients and provide recommendations to address such issues. Medtronic noted that these communications can include letters, emails, calls, press releases, physician notifications and social media postings, as well as informing the FDA and other regulators of the actions.