Alerte De Sécurité sur Trade name: GE T2100 Ergometer Treadmill. Technical Name: Treadmill. ANVISA Registration Number: 80071260189. Hazard Class: I. Affected Model: GE T2100 Ergometer Treadmill. Batch / Affected Series Numbers: GE T2100 Ergometer Treadmill

Selon Agência Nacional de Vigilância Sanitária (ANVISA), ce/cet/cette alerte de sécurité concerne un dispositif en/au/aux/à Brazil qui a été fabriqué par GE Healthcare do Brasil, Com. e Serv. para Equipamentos Médico-Hospitalares Ltda; GE Healthcare Critikon de Mexico S de RL de CV.

Qu'est-ce que c'est?

Les alertes fournissent des informations importantes et des recommandations concernant les dispositifs médicaux. Le fait qu'une alerte soit émise ne signifie pas nécessairement qu'un dispositif soit dangereux. Les alertes de sécurité, qui sont envoyées tant aux travailleurs du secteur médical qu'aux utilisateurs de ces dispositifs, peuvent inclure des rappels. Elles peuvent être rédigées par des fabricants mais aussi par des autorités en charge de la santé.

En savoir plus sur les données ici
  • Type d'événement
    Safety alert
  • ID de l'événement
    2553
  • Date
    2018-04-20
  • Pays de l'événement
  • Source de l'événement
    ANVISA
  • URL de la source de l'événement
  • Notes / Alertes
    Brazilian data is current through June 2018. All of the data comes from Anvisa, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Brazil.
  • Notes supplémentaires dans les données
    For the recommendations given by the company, access the letter to the customer available at: LINK: http://portal.anvisa.gov.br/informacoes-tecnicas13/-/asset_publisher/WvKKx2fhdjM2/content/alerta-2553-tecnovigilancia-ge-healthcare -file-ergometer-ge-t2100-installation-of-parts-reserve-may-result-in-motion-off-the-tape-d / 33868? p_p_auth = bSxZsrle & inheritRedirect; = false & redirect; = http% 3A% 2F% 2Fportal If you want to notify technical and technical complaints, please contact us at the following link: adverse events use the following channels: Notivisa: Notices of adverse events (EA) and technical complaints (QT) for products subject to Sanitary Vigilance should be made through the NOTIVISA System. To access the System, you must register and select the Health Professional option, if you are a liberal professional or the Institution / Entity option, if you are a professional of an institution / entity. Technovigilance System: Patient or citizen can notify through the Technovigilance System / SISTEC access through the link Additional information: - Date of notification notice to Anvisa: 04/18/2018 The company holding the record of the affected product is responsible for contacting its clients in a timely manner to ensure the effectiveness of the ongoing Field Action. It stands out the joint responsibility of the distribution chain and use of health products in maintaining its quality, safety and efficacy, as well as the effectiveness of the Field Action, expressed by RDC 23/2012: "(...) Art 2 ° A holder of registration of health product is the holder of the registration / registration of health product with Anvisa. Single paragraph. The registration holder, as well as other agents involved from the production to the use of the product, or discarding it when applicable, are jointly and severally responsible for maintaining the quality, safety and efficacy of the health products to the final consumer. Art. 12 Distributors of health products shall forward to the registration holder, in a timely manner, the distribution map and other information requested for the notification and execution of field actions. (...) "
  • Cause
    The t2100 treadmill performance problem with customer proprietary spare parts (2026182-002 or 2026182-004) was not solved with the previous safety fix (gehc fmi 30074). if these parts were installed from the customer's reserve stock on the t2100 treadmill, uncontrolled movement of the locomotive strap may occur during the exercise test exercise. this movement may be seen as an unexpected slowdown, followed by a sudden acceleration of the belt in forward or backward direction. when this occurs, even when the stop button is pressed, the movement of the strap does not stop immediately. instead, it will stop in up to 35 seconds by slowing down the maximum speed. sudden change in speed and / or belt direction can result in falls and injuries to the patient.
  • Action
    Field Action Code IMF 30085 triggered under the responsibility of the company GE Healthcare do Brasil, Com. for Equipos Médico-Hospitalares Ltda. Field correction.