Alerte De Sécurité sur Trade Name: SCITECH CATCH TIE. Technical Name: Filters. ANVISA registration number: 10413960200. Risk Class: II. Model affected: NA. Serial numbers affected: 111624 - LACO SNARE SCITECH 4FR 05MM X 120CM KIT. 112000 - LACO SNARE SCITECH 5FR 15MM X 120CM KIT. Lots: 023843, 024004, 024494, 024498, 024895 and 025282

Selon Agência Nacional de Vigilância Sanitária (ANVISA), ce/cet/cette alerte de sécurité concerne un dispositif en/au/aux/à Brazil qui a été fabriqué par Scitech Produtos Médicos Ltda.; Scitech Produtos Médicos Ltda..

Qu'est-ce que c'est?

Les alertes fournissent des informations importantes et des recommandations concernant les dispositifs médicaux. Le fait qu'une alerte soit émise ne signifie pas nécessairement qu'un dispositif soit dangereux. Les alertes de sécurité, qui sont envoyées tant aux travailleurs du secteur médical qu'aux utilisateurs de ces dispositifs, peuvent inclure des rappels. Elles peuvent être rédigées par des fabricants mais aussi par des autorités en charge de la santé.

En savoir plus sur les données ici
  • Type d'événement
    Safety alert
  • ID de l'événement
    2567
  • Date
    2018-05-09
  • Pays de l'événement
  • Source de l'événement
    ANVISA
  • URL de la source de l'événement
  • Notes / Alertes
    Brazilian data is current through June 2018. All of the data comes from Anvisa, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Brazil.
  • Notes supplémentaires dans les données
    Recommendations (what to do, contacts with the manufacturer, how to notify Anvisa, etc.): The company recommends not to use the products under suspicion of diversion and to arrange a return according to Alert message issued to customers who received the diversion lot and other lots under suspicion of diversion. If you wish to notify technical complaints and adverse events use the following channels: Notivisa: Notices of adverse events (EA) and technical complaints (QT) for products subject to Sanitary Surveillance should be made through the NOTIVISA System (http: //portal.anvisa .gov.br / notivisa). To access the System, you must register and select the Health Professional option, if you are a liberal professional or the Institution / Entity option, if you are a professional of an institution / entity. Technovigilance System: Patient or citizen can notify through the Technovigilance System / SISTEC access through the link Complementary Information: - Date of identification of the problem by the company: 06/04/2018 - Date of notification notice to Anvisa: 04/05/2018 The company that owns the affected product is responsible for contacting its customers in a timely manner. to ensure the effectiveness of the Field Action in progress. It stands out the joint responsibility of the distribution chain and use of health products in maintaining its quality, safety and efficacy, as well as the effectiveness of the Field Action, expressed by RDC 23/2012: (...) Art. 2 ° A holder of registration of health product is the holder of the registration / registration of health product with Anvisa. Single paragraph. The registration holder, as well as other agents involved from the production to the use of the product, or discarding it when applicable, are jointly and severally responsible for maintaining the quality, safety and efficacy of the health products to the final consumer. Art. 12 Distributors of health products shall forward to the registration holder, in a timely manner, the distribution map and other information requested for the notification and execution of field actions. (...)
  • Cause
    The company informs that identification of capture loops of 5mm diameter occurred in a batch identified as being 15mm in diameter. after having tracked 2 units consigned in brazil of the claimed lot, it was requested to return it and found that one of the units contained 5mm in diameter, confirming the complaint made by chilean customer. in this way, it was decided to collect all units of products of 5 and 15mm produced with the same batch of raw material.
  • Action
    Field Action Code 074/18, under the responsibility of the company Scitech Produtos Médicos Ltda. Gathering. Product verification and incineration.

Manufacturer