Alerte De Sécurité sur Trade name: SURGICAL TABLES. Technical name: SURGICAL TABLES. ANVISA registration number: 80259110065. Risk class: - Low Risk. Model affected: MOBILE SURGICAL TABLES. Serial numbers affected: All units manufactured in the period between 2009 and 2013.

Selon Agência Nacional de Vigilância Sanitária (ANVISA), ce/cet/cette alerte de sécurité concerne un dispositif en/au/aux/à Brazil qui a été fabriqué par MAQUET DO BRASIL EQUIPAMENTOS MEDICOS LTDA.; MAQUET GmbH.

Qu'est-ce que c'est?

Les alertes fournissent des informations importantes et des recommandations concernant les dispositifs médicaux. Le fait qu'une alerte soit émise ne signifie pas nécessairement qu'un dispositif soit dangereux. Les alertes de sécurité, qui sont envoyées tant aux travailleurs du secteur médical qu'aux utilisateurs de ces dispositifs, peuvent inclure des rappels. Elles peuvent être rédigées par des fabricants mais aussi par des autorités en charge de la santé.

En savoir plus sur les données ici
  • Type d'événement
    Safety alert
  • ID de l'événement
    2053
  • Date
    2016-10-17
  • Pays de l'événement
  • Source de l'événement
    ANVISA
  • URL de la source de l'événement
  • Notes / Alertes
    Brazilian data is current through June 2018. All of the data comes from Anvisa, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Brazil.
  • Notes supplémentaires dans les données
    According to the company, since they can not ensure that other Devices affected by improperly welding have been placed on the market, the component of all possibly affected Devices (1002.65A0) manufactured within the period between 2009 and 2013 will be replaced . In this way, MAQUET's service department or a service technician authorized by MAQUET will contact the customer to arrange a replacement date without charge. The device must be inspected before any use until replacement is performed. In this particular context, special attention should be paid to possible cracks in the weld. According to the company, so far, no incident has been reported in which someone has been injured. It is likely that the break was caused due to improperly welded together with rough handling and multiple application of overload by the user, respectively. ATTENTION: If in doubt, it is advisable that the Device is not used. ### UPDATED ON 10/16/2017, company forwarded evidence and documents proving the correction of the parts and closing the field action. If you wish to notify technical complaints and adverse events use the following channels: Notivisa: Notices of adverse events (AE) and technical complaints (QT) for products subject to Sanitary Surveillance should be made through the NOTIVISA System. To access the System, you must register and select the Health Professional option, if you are a liberal professional or the Institution / Entity option, if you are a professional of an institution / entity. Technovigilance System: Patient or citizen can notify through the Technovigilance System / SISTEC access through the link
  • Cause
    Maquet gmbh cases were reported to have shown that the joint welded between the square structure and the first fixable swivel axis has been broken. this joint must withstand the highest loads due to the cantilever. it is likely that the break was caused due to improperly welded together with rough handling and multiple application of overload by the user, respectively. if the weld is broken during a procedure, there is no avoiding a mechanical injury.
  • Action
    Field Action Code CAPA 2016-008 triggered under the responsibility of the company MAQUET DO BRASIL EQUIPAMENTOS MEDICOS LTDA. Company will make field correction to replace the compromised component.