Alerte De Sécurité sur Wyler AD-TECH Electrode Cable (Cable Connector for Electrode XX Contacts (04, 06, 08, 16, 32 and 64 contacts and Forame Oval 04 contacts) // ANVISA registration number: 10244510006. Technical Name: Cable Connector for Electrode // Hazard Class: IV - Maximum Hazard // Model Affected: L-DCL-4DINX; L-DCL-6DINX; L-DCL-8DINX; L-DCL-16BDINX; L-DCL-32BDINX; L -DLC-64BDINX and FO-LDC-4DINX // Serial numbers affected: 208140559, 208140560, 208140561, 208140562, 208140563 and 208140579

Selon Agência Nacional de Vigilância Sanitária (ANVISA), ce/cet/cette alerte de sécurité concerne un dispositif en/au/aux/à Brazil qui a été fabriqué par Physiomed Importação e Comércio Eireli; AD TECH Medical Instruments Corporation.

Qu'est-ce que c'est?

Les alertes fournissent des informations importantes et des recommandations concernant les dispositifs médicaux. Le fait qu'une alerte soit émise ne signifie pas nécessairement qu'un dispositif soit dangereux. Les alertes de sécurité, qui sont envoyées tant aux travailleurs du secteur médical qu'aux utilisateurs de ces dispositifs, peuvent inclure des rappels. Elles peuvent être rédigées par des fabricants mais aussi par des autorités en charge de la santé.

En savoir plus sur les données ici
  • Type d'événement
    Safety alert
  • ID de l'événement
    1916
  • Date
    2016-04-29
  • Pays de l'événement
  • Source de l'événement
    ANVISA
  • URL de la source de l'événement
  • Notes / Alertes
    Brazilian data is current through June 2018. All of the data comes from Anvisa, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Brazil.
  • Notes supplémentaires dans les données
    Customer Recommendations: AD-TECH believes that these products may be compromised by sterilization, so we urge you to immediately examine your inventory and segregate all products in closed / unused packaging that are the reason for this recall, and return to Physiomed at the following address: Physiomed Importação e Comércio Ltda. If you have knowledge of any patient case that has used the product of the affected lot and has presented adverse reactions, we request that inform us what happened, through the means of contact shown below: Phone for contact: (11) 5593-2240 E-mail: sac@physiomed.com.br / talitha.fernandes@physiomed.com.br. According to the company, to date, technical complaints, occurrences of health damages, adverse reactions or any other complaints related to these products, both in Brazil and internationally, have not been reported to the manufacturer or the holder of the registration in Brazil (Physiomed) . This payment by the manufacturer is a precautionary measure. Customers who received the notified products must check stock and return unused / unopened packaging to Physiomed for later return to the manufacturer.
  • Cause
    During a transport simulation at ad-tech medical instrument corporation, the company concluded that the packaging proved to be fragile during this simulation, which may compromise the sterile barrier of the product.
  • Action
    The manufacturer ADTECH MEDICAL INSTRUMENT CORP. initiated the recall due to the identification of defects in the packaging of these products (which constitutes compromise of the sterile barrier). This was identified during a transport simulation at the Ad-Tech Medical Instrument Corporation because the current packaging of this product is made up of grade-surgical type paper and proved to be brittle during this simulation. Field Action Classification: Return to Manufacturer Return. Field Action Code: 001/2016.