Rappel de MX8000 IDT MULTISLICE C+ IMAGING SYSTEM

Selon Health Canada (via FOI), ce/cet/cette rappel concerne un dispositif en/au/aux/à Canada qui a été fabriqué par PHILIPS ELECTRONICS LTD..

Qu'est-ce que c'est?

Une correction ou un retrait opéré par un fabricant afin de répondre à un problème causé par un dispositif médical. Les rappels surviennent lorsqu'un dispositif médical est défectueux, lorsqu'il pourrait poser un risque pour la santé, ou les deux à la fois.

En savoir plus sur les données ici
  • Type d'événement
    Recall
  • ID de l'événement
    26539
  • Classe de risque de l'événement
    III
  • Date de mise en oeuvre de l'événement
    2013-07-15
  • Pays de l'événement
  • Source de l'événement
    HC
  • Notes / Alertes
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Notes supplémentaires dans les données
  • Cause
    The ct system phantom supplied with the affected scanners has a special pin by which the customer can check the low contrast (lc) detectability for quality assurance. this pin consists of a nylon (aculon) body with six smaller lexan pins of 3 mm 4 mm 5mm 6 mm 7 mm and 8 mm diameters which have contrast difference of approximately 1% from aculon. the lc test is performed by visually inspecting the section and selecting the smallest detectable lexan pin for a predefined scan protocol as defined in the instructions for use (ifu) manual. this manufacturer-recommended procedure is subjective in nature and needs to be performed on several scans and by different people to be reliable. the pin visibility deteriorates over time because of slow water absorption. as the pin visibility deteriorates the test becomes unreliable and inaccurate.

Device

  • Modèle / numéro de série
    Model Catalog: OFF-EI-590 (Lot serial: ); Model Catalog: 362299 (Lot serial: ); Model Catalog: 4535 670 73191 (Lot serial: N/A FOR ALL); Model Catalog: 4535 670 73181 (Lot serial: N/A FOR ALL)
  • Description du dispositif
    MX8000 IDT MULTISLICE C+ IMAGING SYSTEM
  • Manufacturer

Manufacturer

  • Adresse du fabricant
    MARKHAM
  • Société-mère du fabricant (2017)
  • Source
    HC