Alerte Ou Avis De Sécurité sur Procedure packs containing various central venous catheters: pressure injectable, multi-lumen and multi-lumen with Arrowg+ard

Selon Medicines and Healthcare products Regulatory Agency, ce/cet/cette alerte ou avis de sécurité concerne un dispositif en/au/aux/à United Kingdom qui a été fabriqué par Arrow International (a division of Teleflex).

Qu'est-ce que c'est?

Les avis de sécurité sont des communications que les fabricants de dispositifs médicaux ou leurs représentants font parvenir en particulier aux travailleurs du secteur médical, mais aussi aux utilisateurs de ces dispositifs. Les avis de sécurité concernent les actions entreprises quant aux dispositifs présents sur le marché et peuvent inclure des rappels et des alertes.

En savoir plus sur les données ici
  • Type d'événement
    Safety alert / Field Safety Notice
  • ID de l'événement
    MDA/2014/028
  • Numéro de l'événement
    CON435784
  • Date
    2014-07-11
  • Date de publication de l'événement
    2014-07-11
  • Pays de l'événement
  • Source de l'événement
    MHRA
  • URL de la source de l'événement
  • Notes / Alertes
    Data from the United Kingdom is current through April 2019. All of the data comes from the Medicines and Healthcare products Regulatory Agency, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and the United Kingdom.
  • Notes supplémentaires dans les données
  • Cause
    Risk of incorrect lumen selection. the extension lumen may have the wrong exit (proximal, medial, distal) and size printed on it, although the hub colour coding remains correct. for example the distal lumen may be printed with the words ‘medial 18 ga’ instead of ‘distal 16ga’. arrow international issued a second field safety notice (fsn) dated 20 june 2014, which informed users of an extension to their original fsn dated 15 may 2014. this extension now includes affected devices in procedure packs. this medical device alert is an update to mda/2014/022 issued on 12 june 2014. clinicians and nursing staff may rely on the printed information on the catheter lumens to identify them and to select the appropriate lumen to deliver blood and other fluids. misprinted lumens could lead to a range of problems, if used inappropriately.
  • Action
    Identify all additional affected stock using Appendix 2 of the FSN. NHS supply chain codes are detailed below. Ensure that all relevant clinicians and nursing staff are aware of this issue and receive a copy of the Arrow International second FSN. Be guided by Arrow’s hub colour coding, not by what is printed on the lumen. Arrow’s hub colour coding is: White = proximal lumen = 18 GA Grey = medial lumen 1 = 14 GA Blue = medial lumen 2 = 18 GA Brown = distal lumen = 16 GA Report any cases of misprinted lumens to MHRA.

Device

  • Modèle / numéro de série
  • Description du dispositif
    Update to MDA/2014/022 . Procedure packs containing various central venous catheters: pressure injectable, multi-lumen and multi-lumen with Arrowg+ard ® antimicrobial surface. Specific product references and part numbers are affected. Check Appendix 2 of the FSN for the full list of additional affected product codes. If relevant to you, the national supply codes are as follows: England NPC: NHS Supply chain code Manufacturer’s part number Product description FSQ1653 AR12854-MRI Manchester 4L CVC INS Pack 8.5Fr x 16cm FSQ1311 AR22854-HCH HCH 4L AGB CVC PACK 8.5FR 16CM FSQ372 AR22854-INSP INSP 16CM 4L AGB ARS CVC PACK FSQ1654 AR22854-MRI MANCHESTER ROYAL INFIRMARY AGBQUAD Northern Ireland: Manufacturer’s codes as in Appendix 2 of the FSN . Scotland SKU: Manufacturer’s codes as in Appendix 2 of the FSN . Wales NSV: Manufacturer’s codes as in Appendix 2 of the FSN .