Avis De Sécurité sur NC Trek RX Coronaire dilatatiekatheter, NC Traveler RX Coronaire dilatatiekatheter, NC Tenku RX Coronaire dilatatiekatheter,

Selon Inspectie Gezondheidszorg en Jeugd, ce/cet/cette avis de sécurité concerne un dispositif en/au/aux/à Netherlands qui a été fabriqué par Abbott.

Qu'est-ce que c'est?

Les avis de sécurité sont des communications que les fabricants de dispositifs médicaux ou leurs représentants font parvenir en particulier aux travailleurs du secteur médical, mais aussi aux utilisateurs de ces dispositifs. Les avis de sécurité concernent les actions entreprises quant aux dispositifs présents sur le marché et peuvent inclure des rappels et des alertes.

En savoir plus sur les données ici
  • Type d'événement
    Field Safety Notice
  • Date
    2017-03-22
  • Pays de l'événement
  • Source de l'événement
    IGJ
  • URL de la source de l'événement
  • Notes / Alertes
    The data from the Netherlands is current through October 2018. All of the data comes from Inspectie Gezondheidszorg en Jeugd, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and the Netherlands.
  • Notes supplémentaires dans les données
  • Cause
    With products of the identified lot numbers, the protective sleeve may be difficult to remove, which can lead to problems when filling or emptying the balloon. the number of reported cases of balloon filling or emptying problems related to a hard-to-remove sleeve is 0.12%. possible risks associated with this are air embolism, additional intervention, thrombosis, and myocardial infarction.

Manufacturer

  • Société-mère du fabricant (2017)
  • Commentaire du fabricant
    “We are in constant communication with regulatory agencies and competent authorities worldwide which allows us to implement global recalls or in-country communication quickly and effectively,” Abbott, which now owns St. Jude Medical told ICIJ in a statement. In addition to sending global notices to physicians worldwide, we also make sure that product advisories are available online and classification of product recalls and product advisories are determined by global regulatory bodies which can impact the timing in any given country. MD companies follow varying regulations in different countries. In come countries software is not regulated so a recall in one country related to software would not be classified as a recall or field action in another. In addition, review cycles within the regulatory process can be different in each country which can impact communication and recall timing. Note published by the authorities from the Netherlands: This message is a warning from the manufacturer. After placing a medical device on the market, the manufacturer is obliged to follow the device and, where necessary, to improve it. When there is a risk for the user, the manufacturer must take action. The manufacturer informs users and the inspection of this action with a warning (Field Safety Notice). See also medical device warnings. This is part of the supervision of medical technology.""Note published by the authorities from the Netherlands: This message is a warning from the manufacturer. After placing a medical device on the market, the manufacturer is obliged to follow the device and, where necessary, to improve it. When there is a risk for the user, the manufacturer must take action. The manufacturer informs users and the inspection of this action with a warning (Field Safety Notice). See also medical device warnings. This is part of the supervision of medical technology.
  • Source
    IGJ